1-CUBE OXI User Manual - Page 5

Browse online or download pdf User Manual for Measuring Instruments 1-CUBE OXI. 1-CUBE OXI 11 pages. Oxygen measuring device

5.2.2 CHARGING OF THE DEVICE
Fixed type of the device (without battery) has to be permanently connected to the network trough the adapter 24
V DC/200 mA, which is connected with oximeter.
The information below is valid only for the portable type of oximeter supplied with the rechargeable batteries.
If the battery voltage is under 16,5 V low then after the switch-on the device the text „Recharge battery" appears
on the display for a second. It is possible to work with the device for another 30 minutes. If the device is not
recharged during this time then the text „Low voltage" appears on the display and the device is shut down.
If you want to monitor how the battery is charged while working with the device (when it doesn't measure O2) – then
press the rotation key - the text „battery)" appears on the first line of display and current voltage on the second
line of display. Unscrew the connector cover and put the opposite connector of the adaptor in to charge the device.
Warning: wach the right orientation (bump in the adapter connector into the adapter slot. Boss and slot are
marked with coloured dot). Adaptor can be plugged in el. socket 230V, 50Hz now. During charging of the device
the led BAT is shining.
Notice: In case the connectors will not be put in the right way they will be damaged. This type of damage is
not accepted by manufacturer under warranty.
The device can be charged only by the supplied adaptor (output voltage of adapter is 24 V DC/200 mA). Average
charging takes about 8 hrs. I n case of need the period can be shortened but the running period and the accumulator
life-time will be shortened too. Once the charging of the device is over unplug the adaptor and the opposite
connector of the adaptor from the connector on the device. At the end screw fully the connector cover on the device.
5.2.3 O
CONTENT MEASUREMENT
2
Follow the paragraphs 5.1.1. then further:
1) Press the key ON to switch on the device. The text „1-CUBE" with the program version appears on the display for
a moment. Then the following text appears: „ENT.meas."
2) Press the key ENTER. The already measured value of O2 appears on the display - on the first line in ppb,
on the second line in ppm for example: „ppb 1253"
The displayed values are saved in ppb units.
Press the key ENTER to finish the measurement, the text: „ENT.meas."
And the device is ready for next measurement
5.2.4 SCANNING MEMORY OF THE MEASURED VALUES O
The measured values of O2 are saved in the device memory even after its shutdown. The capacity of memory is 450
of measured values.
1) Press the key ON to switch on the device. The text „1-CUBE" with the program version appears on the display for
a moment. Then the following text appears„ENT.meas."
2) Press the key plus (+).There is text „memory" on the first line and the measured values of O
one by one from the oldest value to the newest ones on the second line. After the last saved value the numbers start
to appear
65535 (free capacity of the memory).
following text appears„ENT.meas."
Notice. Once there is the full storage, the memory will be automatically deleted.
Then the following text appears:
„memory"
„deleting"
And all the saved values are overwritten with number 65535 (free capacity of the memory).
„ppm 1,253"
„+ view."
„+ view."
„+ view."
„+ view."
Press the key ENTER to stop scanning memory Then the
2
in
ppb
2
appear