TROUBLESHOOTING (Continued):
b. Verify that the lip clip has a good connection to the
patient's oral mucosa.
c. Verify the file clip metal is clean, free of debris and corro-
sion, clean if not.
d. Verify the pulp chamber is clean and dry.
6. The display indicates only when the file is very near the
apex.
a. The canal may be obstructed by dentin debris or shav-
ings. The unit will only indicate correctly when the file
passes through the apical constriction.
b. The canal may be extremely dry, moisten the canal with
water or sodium hypochlorite.
c. If the apical foramen is too large or incompletely formed,
the indication may suddenly jump as the file tip approach-
es the foramen.
7. There are no audio indications.
a. Verify audio indication has been turned on.
8. The display gives no indications or gives incorrect indi-
cations.
a. Verify that the lip clip has a good connection to the
patient's oral mucosa.
b. Make sure all connections to the cable assembly are
secure. Specifically check the cable assembly jack is
properly inserted in the unit, the lip clip is well seated in
its receiving connector, and the file clip assembly is well
seated in its receiving connector.
c. Verify the file clip metal is clean, free of debris and corro-
sion, clean if not.
d. Verify the pulp chamber is clean and dry.
e. The file clip plastic or metal may be damaged, replace it
if so.
f. Verify the file clip is securely attached to the file and the
file is inserted in the canal.
g. Verify the canal has not been retreated with silver points.
h. Verify the apex is complete.
i. The canal may be calcified.
j. The canal may be obstructed by dentin debris or shav-
ings. The unit will only indicate correctly when the file
passes through the apical constriction.
8.