11. Unplug the firewire cable from the AXP or PS400 camera.
2.1.2 System Start-Up
1. Please see the relevant base TherMoiré system manual for its start-up
procedures.
2. Turn on the Light Source Power switches. Make sure the Light Source
Adjustment knob is placed at approximately 90% of the maximum level.
3. Run the DIC Surface Measurement program by clicking its icon on the Studio
Manager bar
For operation of this program, refer to the Surface Measurement User Manual.
4. Check if there is any dust or other kinds of contaminations on the oven lid glass.
If so, clean the glass according to Appendix A.3.
2.2

System Alignment

1. Choose the largest calibration target that will fit within the cameras' field of view
(FOV). Place the calibration target on the sample support rails such that it is
roughly in the center of the oven. The target should be oriented with the long
dimension running left to right.
2. Close the oven lid, being careful that there is enough of a gap above the
calibration target so that there is clearance when the lid closes.
Figure 2.5 Fully Installed DIC system
Note: If the AXP system camera is connected at
startup, the software may produce an error and not
start.
A start-up splash screen appears while the software initializes.
.