Installation and Maintenance Instructions
for ZT11 High Purity Transmitter
a. Adjust the zero point output in the same way as [(1) Zero
adjustment].
b. Accurately apply pressure upper limit in the pressure range,
and then adjust the output to 20mA DC in span adjustment
potentiometer.
c. Repeat several times alternating Step a and b to make the
span adjustment.
• Do not change zero adjustment while rotat-
ing span adjustment potentiometer.
m
CAUTION:
• Avoid putting an excessive load, >44 in. lb.,
10 sec., on the zero/span adjustment poten-
tiometers as this may result in damage.
• Do not apply pressure above the pressure
range indicated on the device label.
After the adjustment, please be sure to attach the rubber plug
by referring to the figure below.
The "Z" on the rubber plug indicates the position of the poten-
tiometer for zero adjustment, and the "S" indicates the posi-
tion of the potentiometer for span adjustment.
• Ensure rubber plug has been returned to
its original position to retain product pro-
m
tection class integrity. Protection class may
CAUTION:
be impaired if the rubber plug is not firmly
affixed.
10. WARRANTY
If the delivered products within the warranty period (within one
year from the delivered date) are determined to be non-con-
forming products according to "Defects due to the design or
manufacturing", they are repaired or replaced with conforming
products free of charge.
However, note that the following cases are excluded.
(1) Where the delivered products are disassembled, altered or
where their parts are replaced or new function is add by the
customer or any third party
© Ashcroft Inc. 2020, 250 East Main Street, Stratford, CT 06614-5145, USA, Tel: 203-378-8281, Fax: 203-385-0357, www.ashcroft.com
All sales subject to standard terms and conditions of sale. ZT11_Transducer_im_RevA_(I&M011-10274)_04-03-20
(2) Where directions described in the instruction manual or
catalog are not observed.
(3) Where the non-conformance is caused by deterioration due
to use, natural disaster, fire or other force majeure events
(4) The secondary damage caused by non-conformance of the
products including the above.
Returned units are subject to evaluation by Ashcroft prior to
determining warranty.
11. OTHERS
This instruction manual is not intended to cover all equipment
and/or installation and maintenance details. Please contact
Ashcroft if additional detail is sought, or in the event that the
instrument does not adequately serve your company's intentions.
Additionally, this instruction manual may be changed, at
any time, without prior notice for the purposes of version
upgrades, revisions etc.
Thank you for your understanding and cooperation.
7