ASHE STL-94 Operation Manual - Page 13
Browse online or download pdf Operation Manual for Recording Equipment ASHE STL-94. ASHE STL-94 17 pages. Stream selector
Connect a milliamp-meter across 4-20 mA OUTPUT1 terminals with the correct polarity.
The display will show the corresponding ADC count corresponding to 4 mA Output current.
Use the SHIFT key to shift the cursor to desired digit and INC key to change the desired digit from 0 to
9 to get 4 mA on the milliamp meter.
Press MODE key to save the ZERO calibration point of 4-20 mA DC for Output 1.
In the same sequential process, the following parameters for each Stream can be set :
Output Zero Adjust for Streams 2, 3, 4..
High Alarm for Streams 2, 3, 4.
This completes the 4-20 mA retransmission analog output calibration of the instrument.
Press MODE key to exit from Retransmission Calibration Mode. After pressing the MODE key, the LCD
Display will return to normal PROCESS SCREEN.
SETTING MODE
The ASHE Stream Selector Model STL-94 provides both, AUTO and MANUAL modes of operation. One
can select AUTO or MANUAL mode directly from the front panel key. There are two sub-mode options
in the AUTO mode of operation, viz. Timer and Pulse modes.
Enter the number 06 into the Select Menu by using INC key in the PARAMETER SETTING MODE
SCREEN.
Press MODE key to enter into the Setting Mode.
The LCD Display shows :
Press the SHIFT key to move the cursor to the last digit. Enter Password 0004 by using INC key and
press MODE key. After pressing the MODE key, the LCD Display shows
Use the INC key to select the desired mode of operation to either Timer or Pulse.
Press the MODE key to proceed further. After pressing MODE key, the microcontroller will save
selected mode of operation and move to Sample Time setting.
Password
0000
Select Mode
Timer
ASHE CONTROLS PRIVATE LIMITED.
S
S
STL-94
TREAM
ELECTOR
Operation Manual