2. SPECIFICATIONS
2.1.
PERFORMANCE
Dynamic Range 100:1.
Pulse-Pair Resolving Time Output pulse width plus
Delay (as selected by the front panel Delay control), plus
100 ns for negative NIM output or plus 200 ns for positive
NIM output. Minimum resolving time for negative output,
220 ns; for positive output, 800 ns.
Threshold Temperature instability <0.015%/° C of full
scale, 0 to 50°C using a NIM class A power supply
(referenced to -12 V).
Discriminator Noniinearity <±0.25% of full scale (inte
gral) for LL discriminator.
Delay Temperature instability <0.04%/°C of full scale,
0 to 50° C.
Delay Noniinearity <±2%.
Window Width Constancy <0.1% variation of full-scale
window width over the linear 0- to 10-V range.
Minimum input Threshold 50 mV for Lower Level dis
criminator.
Time Shift vs Pulse Height (Walk)
Walk (ns)
Dynamic
Range
10:1
50:1
100:1
System
System
A
B
±1.0
±2.5
±3.0
±2.0
±4.0
±8.0
System A: Using an ORTEC 460 Amplifier, single delay-line mode,
integrate ^0.1 ms with delay line 1 mS-
System B: Using an ORTEC 471 or 472A Amplifier, unipolar output with
0.5 fiS shaping time, input from ORTEC 419 Pulser.
2.2.
CONTROLS
Lower Level
Front panel 10-turn potentiometer adjust
able from 0 to 10 V; when the rear panel LL Ref mode
switch is set on Int, determines the threshold setting for
the Lower Level discriminator. When the LL Ref mode
switch on the rear panel is in the Ext position, this control
is ineffective.
Window or Upper Level
Front panel 1-turn potenti
ometer determines the window width (0 to +1 V) in the
Window mode or the Upper Level (0 to +10 V) threshold
in the Normal mode. This control is disabled in the
Integral mode.
Int/Nor/Win
Front panel 3-position locking toggle
switch selects one of three operating modes.
Integral LL sets a single-discriminator threshold (0
to +10 V) and UL response is ignored for SCA outputs.
Normal UL and LL are independently adjustable lev
els (0 to +10 V); differential mode operation.
Window
LL sets the baseline level (0 to +10 V) and UL
sets the window width (0 to +1 V); differential mode
operation.
Delay
Front panel 1-turn potentiometer for continuous
adjustment of output delay over a 0.1 to 1.1 fis range.
Walk Adjust Front panel screwdriver adjustment for
precise setting of walk compensation.
LL Ref Mode
Rear panel two-position locking toggle
switch selects either the front panel LL potentiometer or
the voltage signal applied to the rear panel LL Ref Ext
connector as the LL discriminator reference threshold.
2.3.
INPUTS
Signal Input Front panel dc-coupled BNC connector
accepts positive unipolar or bipolar signal, 0 to +10 V
linear range, ±12 V maximum; width, ^100 ns; lOOOfl
input impedance.
LL Ref Ext When the rear panel LL Ref mode switch is
on Ext, the rear panel LL Ref Ext BNC connector accepts
the Lower Level biasing (an input of 0 to -10 V on this
connector corresponds to a range of 0 to 10 V for the
Lower Level discriminator setting). Input protected to
±24 V.
2.4.
OUTPUTS
Pos Out Front panel BNC connector provides positive
NIM-standard output, nominally +5 V; <20 ns rise time,
500 ns wide; <10n output impedance. Output occurs at
the mid-point of the linear input trailing edge plus the
output Delay as selected by the front panel control.
Neg Out Front panel BNC connector provides negative
NIM-standard output, nominally -16 mA (800 mV on 50fi
load); width, <20 ns; rise time, <5 ns. Output occurs at
the mid-point of the linear input trailing edge plus the
output Delay as selected by the front panel control.
UL Out Rear panel BNC connector provides NIM-
standard output, nominally +5 V, 500 ns wide; <10n
output impedance. Output occurs as leading edge of
linear input crosses the UL threshold.
2.5.
RELATED EQUIPMENT
The 553 is compatible with all ORTEC amplifiers and
other amplifiers having a 0- to 10-V positive, linear output
range.
2.6.
ELECTRICAL AND MECHANICAL
Power Required
+24 V, 82 mA; +12 V, 160 mA;
-24 V, 50 mA; -12 V, 120 mA.
Dimensions NIM-standard single-width module (1.35
by 8.714 in.) per TID-20893 (Rev).