EG&G ORTEC 416 Instruction Manual - Page 13

Browse online or download pdf Instruction Manual for Portable Generator EG&G ORTEC 416. EG&G ORTEC 416 15 pages. Gate and delay generator

7 - 1
Table 1, DC Voifoge Measurements
Checkpoint
Average Values
+ 12 bus
+11.8
+ 24 bus
+23.8
-24 bus
-23.6
-12 bus
-11.7
Q2-B
+ 0.70
Q2-C
+ 5.0
Q3-B
+11.5
Q4-B
+21.5
Q5-B
- 0.60
Q6-C
-11.8
Q8-B
+ 0.60
Q8-C Amp Pot Min
+9.4
Q8-C Amp Pot Max
+ 0.50
Q9-E
+12.4
QID-B
- 0.25
QIO-E
+ 0.50
*Q10-C
- 0,45
Qll-E
+ 0.1
Q12-E
- 0.1
Q13-B
+ 0.1
Q14-B
'
+ 1.8
Q15-E
+12.5
* R 33 to be selected, if necessary. Nominal 22 Q
Note: All voltages are read with respect to ground with a vtvm,
with no signal input, and with all front panel pots in the