Askle Sante Winncare Automorpho Plus User Manual - Page 11
Browse online or download pdf User Manual for Medical Equipment Askle Sante Winncare Automorpho Plus. Askle Sante Winncare Automorpho Plus 20 pages.
U
-
N
INSTALLING THE SUPPORT
1.
Disconnect the compressor from the mains
2.
Open the CPR valve to deflate the support. Complete removal of air from the
support can be facilitated by using the inflation/deflation kit connected to the
CPR, and obstructing the power supply pipe.,
3.
Position the compressor in the middle of the support.
4.
Roll the support starting with the feet
5.
Strap up the support
6.
Insert the support in the transport bag supplied with the product.
7.
Position the dirty/clean label in order to make the yellow side visible.
Blue label: the support is clean
Use of blue transport bag
WARNING:
Accidental deflation of the mattress (power cut or accidental cut of the air circuit):
Should the patient have to be kept in the continuous seated position and/or the
patient is unable to get up alone or assisted and/or the patient lives alone and/or
the patient is hard of hearing, the patient must be lifted to replace the mattress of
the medical bed as soon as possible. Should the patient be able to get up alone
and intervention is possible within less than 8 hours, he/she must lower the chest relief
at least.
S
TORAGE AND TRANSPORT
There is a label on the transport bag to enable rapid identification of the device from
the AXTAIR range contained in the bag. This label must be visible when the product is
stored.
The following storage and transport conditions must be followed:
Temperature:
Hygrometry level:
Atmospheric pressure: between 50 kPa and 106 kPa
WINNCARE FRANCE
Yellow label: the support is dirty (risk of
Use of red transport bag (sold as an option)
between -25°C and + 70°C
between 30% and 95%
- 10 -
contamination)