Ambu 624001000US Gebruiksaanwijzing - Pagina 13
Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} Ambu 624001000US. Ambu 624001000US 17 pagina's.
Do not open the tip of the endotherapy instrument or extend the tip of the endotherapy
instrument from its sheath while the instrument is in the working channel, as this may damage
both the endotherapy instrument and the endoscope.
Insertion of active endotherapy instruments 7d
Use of active endotherapy instruments should always be operated according to the respective
manufacturer's instructions for use. Users shall always be familiar with safety precautions and
guidelines on the proper use of active endotherapy instruments, including use of adequate
personal protective equipment.
Do not activate an endotherapy instrument in the working channel before the instrument
distal end can be seen in the image.
It should be recognized that the use of active endotherapy instruments may interfere with
the normal endoscopic image and this interference is not indicative of a malfunction of
the endoscopic system. A variety of factors can affect the quality of the endoscopic image
during use of active endotherapy instruments. Factors such as intensity, high power setting,
close distance of the instrument probe to the endoscope tip and excessive tissue burning
can each adversely influence image quality.
Withdrawal of the endoscope 8
When withdrawing the endoscope, make sure that the control lever is in the neutral position.
Slowly withdraw the endoscope while watching the live endoscopic image.
4.3. After use
Visual check 9
1. Are there any missing parts on the bending section, lens, or insertion cord? If yes, then take
corrective action to locate the missing part(s).
2. Is there any evidence of damage on the bending section, lens, or insertion cord? If yes,
then examine the integrity of the product and conclude if there are any missing parts.
3. Are there cuts, holes, sagging, swelling or other irregularities on the bending section,
lens, or insertion cord? If yes, then examine the product to conclude if there are any
missing parts.
In case of corrective actions needed (step 1 to 3) act according to local hospital procedures.
Disconnect
Disconnect the endoscope from the displaying unit 10 . The aScope 5 Broncho is a single-use
device. Do not soak rinse, or sterilize this device as these procedures may leave harmful
residues or cause malfunction of the device. The design and materials used are not compatible
with conventional cleaning and sterilization procedures.
Disposal 11
The used aScope 5 Broncho is considered contaminated after use and must be disposed
of in accordance with local guidelines for collection of infected medical devices with
electronic components.
5. Technical product specifications
5.1. Standards applied
The endoscope conforms with:
– EN 60601-1 Medical electrical equipment – Part 1: General requirements for basic safety and
essential performance.
– EN 60601-2-18 Medical electrical equipment – Part 2-18: Particular requirements for the basic
safety and essential performance of endoscopic equipment.
– IEC 60601-1-2 Medical electrical equipment – Part 1-2: General requirements for basic safety
and essential performance – Collateral standard: Electromagnetic disturbances –
requirements and tests.
13