Flomec TM30 Series Handleiding voor producteigenaren - Pagina 9
Blader online of download pdf Handleiding voor producteigenaren voor {categorie_naam} Flomec TM30 Series. Flomec TM30 Series 13 pagina's. Electronic water flowmeters 3 inch & 4 inch meter sizes

OPERATION / CALIBRATION
Verify Meter Accuracy
1.800.561.8187
Before using, check the meter's accuracy and verify calibration.
1. Make sure there is no air in the system by starting the flow until it runs
steadily. Then, stop or divert the flow using a valve or nozzle.
2. Meter an exact known volume into an accurate container. For best
results, meter with one continuous full stream.
3. Check the volume against the display or recording equipment. If the
amount metered is accurate, further calibration is not necessary. If not,
refer to the Q9 Owner's Manual (Non-Agency) Field Calibration Section
for further instructions (see below).
NOTE: Make sure you meet the meter's minimum flowrate requirements:
TM Series Meter Minimum Flowrate
TM30 - 3 inch
TM40 - 4 inch
You can download the Q9 Owner's Manual (Non-Agency) here:
www.
30 GPM (113.6 L/min)
40 GPM (151.4 L/min)
.com
9