STEP 4: To close the cover, push down sharply on arrow at front of cover with the ball of your foot.
Corded Connection Option for the FOOT CONTROL
Your wireless foot control has a corded connection option. This will allow you to use the foot control as a wired foot control if
preferred. To use the corded connection, connect the optional foot control cable to the F.C. Port on the back of the foot control and to
the back of your laser on the receptacle marked F.C. Remove the battery for wired operation.
Corded Connection
CORDLESS FOOT CONTROL BATTERY REPLACEMENT
When the red indicator light is blinking on the cordless foot-control, the batteries are low and need to be replaced. Turn off unit, and
switch key to off position. Remove the battery cover and remove the batteries. Discard the batteries as required by applicable laws
and regulations. Replace with a new alkaline AA size battery as indicated in the battery compartment. Replace battery cover and
verify operation by depressing the foot-control button, returning the unit to standby, and viewing the green activity light.
WARNINGS
DANGER: DO NOT USE IN THE PRESENCE OF FLAMMABLE ANESTHETIC MIXTURE WITH AIR, OXYGEN, OR NITROUS OXIDE.
DO NOT RE-USE DISPOSABLE OR SINGLE-USE ITEMS. FAILURE TO COMPLY MAY RESULT IN REINFECTION OR CROSS-INFECTION AND IS A
HAZARD TO THE PATIENT.
DO NOT USE THIS UNIT IF YOU SUSPECT IT IS FUNCTIONING IMPROPERLY OR OTHER THAN DESCRIBED HEREIN.
Eyewear
Post appropriate warnings for laser in use in operatories where the laser is to be used so that appropriate eyewear can be donned
before the entry into the operatory. The doctor, patient, assistant, and any other persons in the operatory during laser procedures
must wear the appropriate laser eyewear protection for the diode laser wavelength of 810 ± 10 nm with an optical density (OD)
of at least +5. Class IV laser radiation is hazardous to the eye from the direct beam and diffuse reflections. It also represents a
significant skin and fire hazard. Safety eyewear not designed to this specification is not suitable for use with Picasso lasers. Laser
protective eyewear is available from your AMD LASERS representative.
Anesthesia
If anesthesia is required for treatment, closely monitor patients for signs of pain or discomfort. If such signs are present, adjust
settings, apply anesthesia, or cease treatment, if necessary.
Adjacent Structures
Picasso and Picasso Lite are designed to remove soft tissue. Therefore, always be aware of adjacent structures and substructures
during treatment and monitor them closely. Be extremely careful not to penetrate or ablate underlying or adjacent tissues. Do not
direct laser energy toward hard tissues, such as tooth or bone. Directing laser toward amalgam, gold, metallic surfaces or reflective
surfaces can result in laser beam reflection and/or scattered radiation.
Suction
Use high-speed suction as required to maintain a clear field of vision during treatment. Do not use Picasso and Picasso Lite if you
cannot clearly see the treatment site. It is recommended that a smoke evacuator or in-line filter be used to capture the plume during
Insert foot control cable in
connection labeled F.C.
Page 16
the treatment of herpetic lesions, oral papillomas, and whenever possible. It is recommended that personnel wear a half face
respirator/clinical mask with proper HEPA™ filtration. The plume should be regarded as a source of active biological material as it
may contain viable tissue particulates.
Training
Only licensed professionals who are certified in Class 4 laser use and have read and understood this Operating Instructions Manual
shall use this device. AMD LASERS
®
assumes no responsibility for parameters, techniques, methods, treatments, or results. Clinicians
and/or operators must use their own clinical judgment and professionalism in determining all aspects of treatment, technique, proper
power settings, interval, duration, etc. A six hour laser certification course is available from AMD LASERS prior to use.
Heat Generation
The end of the fiber optic cable, disposable tips, and non-disposable tips, cannulas, and the handpieces may become hot during use.
Avoid unnecessary contact with tissue.
Disposable Tips
The disposable tips are designed and labeled for SINGLE USE only. These tips are not meant to be bent more than one time, and
bent no more than 90°. They are designed to be bent once to the preferred angle, used on a single patient, and then disposed of
properly. Bending the disposable tip more than once in an unintended manner is a risk to the fiber optic cable inside the cannula
leading to potentially lowered power output. When the handpiece is not in use, place protective cap over the tip to prevent dust or
contamination.
Additionally, bending the metal cannula in the Disposable Tip in an unintended manner can lead to a breakage resulting in a
potential sharp point which could puncture the skin of the patient or office personnel. Care should be taken to not bend the metal
cannula multiple times.
CAUTION: ALWAYS CHECK FOR PRESENCE OF THE AIMING BEAM BY SHINING THE TIP ONTO A NON-REFLECTIVE SURFACE BEFORE FIRING THE
LASER. IF THE AIMING BEAM SETTING ON THE DEVICE IS NOT "OFF" AND YOU DO NOT SEE THE AIMING BEAM, DO NOT USE THE TIP. IT MAY
BE BROKEN WHICH COULD RESULT IN HEATING OF THE METAL CANNULA POTENTIALLY CAUSING AN ACCIDENTAL BURN TO THE PATIENT.
PRECAUTIONS
AMD Picasso and Picasso Lite Diode Laser Systems are for Dental Use Only according to the Indications for Use in this OIM Version
7.1 of the manual. The laser is to be used under the supervision of a Dental professional.
Laser tips, once initiated, are extremely HOT. Touching flammable surfaces with an initiated tip is not recommended.
Periodic cleaning of the tip during procedures is required to keep the efficiency of the heat transfer. If removing coagulum, use
CaviCide™ or other equivalent solution following manufacturer's instructions. If using any product containing alcohol, make sure to
allow tip to air dry fully prior to re-initiating the tip.
When placing the handpiece on the laser holder, make sure to have the tip pointing away from all plastic articles to avoid damage to
the plastic and/or the tip.
Units can only accept one input at a time. Laser settings will reflect the last input from the user.
CAUTION: DIODE LASERS ARE ATTRACTED TO MELANIN AND HEMOGLOBIN. POWER LEVELS ARE INVERSELY RELATED TO THE AMOUNT OF
PIGMENTATION AND VASCULARITY OF THE TARGETED TISSUE. AS PIGMENTATION VARIES FROM PATIENT TO PATIENT, START WITH THE LOWEST
POWER SETTING AND INCREASE AS NEEDED TO ACCOMPLISH THE PROCEDURE IN TISSUES WHICH HAVE LOWER MELANIN AND VASCULAR
CONTENT; USE ANESTHETIC AS NEEDED FOR PATIENT COMFORT.
STEP-BY-STEP INSTRUCTIONS
MULTI-TIP HANDPIECE WITH FIBER
The delivery system use in conjunction with removable and disposable tips consists of the Multi-Tip Handpiece/fiber assembly:
Multi-Tip Handpiece & Fiber
• Picasso (orange) / Picasso Lite (black)
• Length: 2 meters
• F.C. Connector and Handpiece
NOTE: Tips are not autoclavable.
FIGURE 11: FIBER SPOOL
FIGURE 12: MULTI-TIP HANDPIECE & FIBER