Guidance and manufacturer's declaration – electromagnetic immunity
The Ameda Elite Breast Pump is intended for use in
the electromagnetic environment specified below.
I
test
IEC 60601 test
MMUNITY
level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
applies.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for
radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnet-
ic environment due to fixed RF transmitters, an electromagnetic site
The Ameda Elite Breast Pump is intended for use
in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of
the Ameda Elite Breast Pump can help prevent
electromagnetic interference by maintaining a minimum
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not
listed above, the recommended separation distance d in
maximum output power rating of the transmitter in
Compliance level
3 Vrms
3 V/m
Recommended separation distances between portable and mobile RF
communications equipment and the Ameda Elite Breast Pump
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1,2√P
0,12
0,38
1,2
3,8
12
P is the
The customer or the user of the Ameda Elite Breast Pump
should assure that it is used in such an environment.
Electromagnetic environment – guidance
Portable and mobile RF communications
be used no closer to any part of the Ameda Elite Breast
Pump, including cables, than the recommended
Recommended separation distance
d = 1,2√P
d = 1,2√P 80 MHz to 800 MHz
d = 2,3√P 800 MHz to 2,5 GHz
Where P is the maximum output power rating of the
manufacturer and d is the recommended separation
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey,
marked with the following symbol:
survey should be considered. If the measured field strength in the
location in which the Ameda Elite Breast Pump is used exceeds the
applicable RF compliance level above, the Ameda Elite Breast Pump
should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary,
such as re-orienting or relocating the Ameda Elite Breast Pump.
b
should be less than 3 V/m.
distance between portable and mobile RF communications
as recommended below, according to the maximum output
m
80 MHz to 800 MHz
d = 1,2√P
0,12
0,38
1,2
3,8
12
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
should be less than
b
800 MHz to 2,5 GHz
d = 2,3√P
0,23
0,73
2,3
7,3
23
9