Aktakom ATK-2011 Gebruikershandleiding - Pagina 7

Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Aktakom ATK-2011. Aktakom ATK-2011 11 pagina's. Clamp meter

Aktakom ATK-2011 Gebruikershandleiding
c. Read the measured value from the LCD display. The actually measured frequency will be displayed in the upper
LCD.
NOTE: If the true RMS value is less than 3A, the frequency of the signal will be displayed as 0 (Please refer to the
following figure.

III.1.2. HOLD, MAX, MIN and PEAK of AC Current.

a. Set the sliding switch at A.
b. Connect the flexible probe around the conductor. Keep the probe coupling more than 25mm away from the
conductor.
c. The measured value from the LCD display. If the HOLD button is pressed, the symbol of "HOLD", "MAX", "MIN"
or "PEAK" will be shown in LCD alternatively. And the value of the HOLD, MAX, MIN or PEAK function will be
displayed in LCD alternatively.
d. To return to the display of current measurement, hold the HOLD button for more than 2 seconds.
NOTE: The PEAK function displays the maximum value of the input waveform. The sampling time for the PEAK
function is 39 μs (50Hz) or 33μs (60Hz). The HOLD, MAX, or MIN function displays the true RMS value.

III.2 Measurement of ACV

NOTE: If the peak value of the input AC voltage is greater than the maximum value of the range, then symbol of OL
will be displayed.
WARNING: Maximum input for ACV is 600. Do not attempt to take any voltage measurement that exceeds the
limits. Exceeding the limits could cause electrical shock and damage to the clamp meter.
www.tmatlantic.com
5