Daavlin 1 Series Bedieningshandleiding - Pagina 17

Blader online of download pdf Bedieningshandleiding voor {categorie_naam} Daavlin 1 Series. Daavlin 1 Series 19 pagina's. Modes of operation - timed mode

The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment.
IEC 60601
Immunity test
test level
a. Field strength from fixed transmitters, such as base stations for radio (cellular / cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the device is used exceeds
the applicable RF compliance level above, the device should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such
as re-orienting or relocating the device. Over the frequency range 150 kHz to 80 MHz, field
strengths should be less than 3V/m
MNL-00049 [1]
Compliance
level
Proprietary and Confidential. Printed copies will not be updated.
Electromagnetic environment - guidance
Page 17 of 19