Daavlin DT Home-Control Bedieningshandleiding - Pagina 5
Blader online of download pdf Bedieningshandleiding voor {categorie_naam} Daavlin DT Home-Control. Daavlin DT Home-Control 19 pagina's. Modes of operation - timed mode
7.0 Precautions and Warnings
•
To protect the eyes during operation, the operator and anyone in view of the device must wear
the provided UV blocking glasses or goggles designed to block 100% of all UVA and UVB
light from the eye area when worn. Always use Daavlin approved eyewear.
•
Do not remove protective eyewear, or any other protective equipment, during treatment.
•
If psoralens (photosensitizing drugs) are being used as part of your treatment ("PUVA"), your
eyes should be protected from exposure to ultraviolet (sunlight) for 24 hours after taking the
drug. Ultraviolet blocking glasses are provided with devices equipped with UVA lamps.
•
Do not use over skin eruptions without express consent from the attending physician.
•
If a patient experiences burning, never treat the patient until the noticeable effects of burning
subside, and always reduce the subsequent treatment time.
•
To protect unaffected skin during operation, the operator, patient, and anyone in view of the
device must generously apply UV blocking skin protection to all exposed skin that the
physician does not intend to treat.
•
During extremely long treatments, some surfaces may become hot. Remove any body parts
from those surfaces if it becomes uncomfortable. The unit should have a 50% duty cycle of
10 minutes on, 10 minutes off.
•
Erythema can result in as little as 15 seconds of exposure, or approximately 200 millijoules of
dose to UVB light. Prior to using your home phototherapy unit, contact your prescribing
physician for specific treatment instructions and dosing information.
•
Center patients between the lamps during treatment to avoid over exposure to isolated areas
of the body.
•
All people and pets should leave the treatment area to avoid exposure to ultraviolet light.
•
Ask your doctor about protecting areas of the body that have not been exposed to sunlight.
•
Do not use this device for anything other than its intended purpose.
•
This device is only to be used by authorized users.
•
Do not operate this device with a damaged cord or plug.
•
If appliance coupler or mains plug is used as the mains disconnect, do not position the ME
equipment so that it is difficult to operate the disconnection device.
•
To avoid the risk of electric shock, this equipment must only be connected to a supply main
with protective earth.
•
To prevent electric shock, remove power to the device prior to cleaning and servicing.
MNL-00049 [1]
Proprietary and Confidential. Printed copies will not be updated.
Page 5 of 19