In case curing needs to be repeated, press the
Ultraled activation button again. In order to
interrupt the curing cycle, press the activation
button again.
Bleaching
Attach the acrylic bleaching tip (fig. B, item 3 or 4)
to the front orifice of the handpiece.
The bleaching tips are sold separately.
After the tip has been attached to the
equipment, it should be cleaned with some
gauze, since oiliness from the user's hands
and other stains on the acrylic tip may affect
light conduction.
Plug the equipment in the power source outlet, and
turn on its master switch.
Follow the gel manufacturer's instructions as regards
activation time and the distance between the tip
and the gel.
Press the button to activate the LED, which will light
up slowly. As soon as it is activated, the LED will
remain on during the set time and a tone will be
heard every 10 seconds.
When the programmed time has ended, the LED will
go off. (refer to "Time setting").
To interrupt the bleaching cycle, press the
activation button again.
Time Setting
Activate the master switch to turn on the instrument
and keep the handpiece activation button pressed
until a long tone is heard. Release the activation
button. Press it the number of times corresponding
to the desired tome setting. Each touch corresponds
to 10 seconds. The maximum setting time is 100
seconds.
Intensity time adjustment program
Press the activation button on the handpiece
to activate the LED. Then, without releasing
the activation button, wait three seconds, and
a long tone will be emitted. Still keeping the
activation button pressed, wait until a second
tone is emitted. Release the activation button and
wait for a tone indicating the first setting phase,
"Low Power". Then, press the activation button
to set the application percentage. Each touch
corresponds to 10% of the total pre-adjusted total
functioning time, and the maximum will be 100%
(in case the activation button is not pressed, the
system will understand that no percentage has
been set in this phase).
After the desired application percentage has been
programmed, and if it is less than 100%, wait 2
seconds, and 2 tones will be emitted, indicating
8
the second setting phase, "Medium Power".
Repeat the procedure used for the first phase. At
this moment, only the remaining percentage can
be programmed. "High Power" will correspond
to the remaining percentage from the two first
phases. Two seconds after the setting of the
second phase has been concluded, 3 fast tones will
be emitted, indicating setting completion. In case
you wish to set only some of the phases, access
the setting model as explained above, but do not
press the activation button after each phase. The
system will automatically shift to the next phase.
When the activation button is not pressed during
the setting of phase 1 and 2, the application will
be set at 100% with "High Power".
Example: For an application time setting of 60
seconds and the phases respectively set at 10, 20
and 70%, the LED will remain on for 6 seconds at
low power, for 12 seconds at medium power and
for 42 seconds at high power.
Cleaning
Cleaning
Cleaning
Clean the equipment surface with a clean soft
cloth moistened in Aplic Odonto by Dabi Atlante or
another product with similar chemical properties
The use of other chemical products is not
recommended as it may damage the equipment.
In case PVC film is used, make sure that the
ventilation orifices (air inlet and outlet) are not
clogged. The acrylic light tip should never be
wrapped in any kind of material.
Sterilization
Sterilization
Sterilization
The acrylic curing light tip as well as the bleaching
light tips must not be autoclaved.
Maintenance
Maintenance
Maintenance
Preventive maintenance
In order to reduce equipment faults and increase
durability, call Dabi Atlante Authorized Technical
Assistance and set a regular preventive maintenance
plan.
Corrective maintenance
In case the equipment presents any faults that are
not listed in "Troubleshooting", contact Dabi Atlante
Authorized Technical Assistance.
Do not open the equipment and/or try to repair
it as such action may aggravate the problem or
even generate further malfunctioning.
Circuit diagrams, list of parts, components,
others
In case they are necessary, ask Dabi Atlante for
electric diagrams and/or lists of parts, components