dacell DN-AM210 Gebruikershandleiding - Pagina 5

Blader online of download pdf Gebruikershandleiding voor {categorie_naam} dacell DN-AM210. dacell DN-AM210 14 pagina's. Dynamic strain amplifier

⑫ Sensitivity adjustment (GAIN)
X1, X10,X100,X400, X1000 within  1% at error each point
(SPAN)
(FINE)
⑬ S/N rate (Input conversion numerical value): more than 54dB (1kHz)
CMRR (Input conversion numerical value): more than100dB (1kHz)
⑭ Response frequency characteristic
DC~20kHz (GAIN : 1000, Reduction : -3dB)
DC~10kHz (GAIN : 1000, Reduction : -0.1dB)
⑮ Low Pass Filter
(L.P.F) Frequency : 10, 100, 1K, 10K, WB 5 stages
About decrease : within -6dB/oct
Temperature range for use
16
0 ~ +60C
Effect by temperature change
17
Zero drift : within  1 strain/C
Sensitivity change:  0.03% F.S/C
Effect by electric power change (About  10% change in supply voltage)
18
Zero drift: within  0.05% F.S
Sensitivity change: within  0.05% F.S
Supply Voltage
19
AC100/220V  10% 60Hz
External Size and Weight
20
128.4 (H) × 43.2 (W) × 198 (D) mm
800g
OPTION
21
Output: OUT1  100mA output
X1 ~ X11 continuous variableness
X1 ~ X2 continuous variableness
4