Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Astrea Evolve. Astrea Evolve 19 pagina's. 350mm column instructions for replacing the wetted flowpath using the column refresh kit
STAGE 4 - Column Reassembly and
Preparation For Use
Note: Reassembly of the column is the reverse of disassembly and starts from column bottom upwards. Each sub
assembly is reassembled before attached to the column.
The column is supplied with alignment indicators , these are attached to the top flange, tube flanges, bottom flange
and base plate. These assist with the reassembly of the column and ensure correct orientation of components.
Alignment Indicators
Figure 15 Alignment Indicators on Evolve® 350 column
®