DotFit Exerspy Gebruikershandleiding - Pagina 15

Blader online of download pdf Gebruikershandleiding voor {categorie_naam} DotFit Exerspy. DotFit Exerspy 20 pagina's. Activity and calorie tracking system
Ook voor DotFit Exerspy: Gebruikershandleiding (20 pagina's)

Guidance and Manufacturer's Declaration - Immunity
The display is intended for use in the electromagnetic environment specified below.
The customer or user of the display should ensure that it is used in such an
environment.
Immunity Test
ESD
±6kV Contact
IEC 61000-4-2
±8kV Air
EFT
±2kV Mains
IEC 61000-4-4
±1kV I/Os
Surge
±1kV Differential
IEC 61000-4-5
±2kV Common
Voltage Dips/
>95% Dip for
Dropout
0.5 Cycles
IEC 61000-4-11
60% Dip for 5 Cycles
30% Dip for 25 Cycles
Power Frequency
3A/m
50/60Hz
Magnetic Field
IEC 61000-4-8
User Guide
IEC 60601
Compliance
Test Level
Level
±6kV Contact
±8kV Air
N/A
N/A
N/A
3A/m
Electromagnetic
Environment - Guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are synthetic, the r/h
should be at least 30%.
N/A
N/A
N/A
Power frequency magnetic
fields should be that of a
typical commercial or hospi-
tal environment.
14