Com-Power LI-3100 Gebruiksaanwijzing - Pagina 5
Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} Com-Power LI-3100. Com-Power LI-3100 11 pagina's. Line impedance stabilization network 10 khz to 400 mhz

3.0 Product Specifications
Model:
Electrical
Frequency Range:
Compliant Standards:
Number of lines:
Max current rating:
Max Voltage Rating:
Insertion Loss:
Isolation:
Inductor type:
EUT & Power Connector:
RF measurement connector:
Cooling:
Power Adapter for fan:
Fuse:
Adapter output connector Diameter:
Operating Temperature:
Mechanical
Dimensions of each enclosure (L x W x H):
Weight (each network):
This equipment is designed for indoor use only.
3.1
Other items available from Com-Power
Distributed by: Reliant EMC LLC, +1 408 916‐5750, [email protected], www.reliantemc.com
SPA-series Spectrum Analyzers
LIN-120A, LISN 150 kHz – 30 MHz, 20 A
LI-220A, LISN
10 kHz – 30 MHz, 20 A
LI-125A, LISN
150 kHz – 30 MHz, 25 A
LI-550A, LISN
150 kHz – 30 MHz, 50 A
Special order LISNs with higher voltage and / or current
LIT-153 Transient Limiter
Cables and connectors
INSTRUCTION MANUAL FOR LI-3100
LINE IMPEDANCE STABILIZATION NETWORK (LISN)
LI-3100
10 kHz to 400 MHz
RTCA DO-160 / MIL-STD 461 / CISPR 25 / CISPR 16-1-2
Two lines
100 Amp AC, 70 Amp DC
480VAC, 50/60 Hz (Line to Ground) / 676 VDC
<0.65 dB (100 kHz – 108 MHz)
>35 dB (100 kHz – 200 kHz)
>40 dB (200 kHz – 108 MHz)
5 µH
100 Amps Shrouded Superior Electric sockets
50Ω, N type (female)
Forced air
Output: 12 VDC,1 Amp
Input: 120 VAC, 60 Hz / 230 VAC, 50 Hz
1 Amp, 250 V (T)
2.1 x 5.5 mm Center Positive
5° C to 40° C / 40° F to 104° F
10 x 10 x 21 inches / 25.4 x 25.4 x 53.3 cm
13 lb. / 5.9 kg
Page 5 of 11