Dover HYDRO HYDSPD0091 Gebruikershandleiding - Pagina 17

Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Dover HYDRO HYDSPD0091. Dover HYDRO HYDSPD0091 20 pagina's. With evoclean

Dover HYDRO HYDSPD0091 Gebruikershandleiding
3.OO programming (continued)
3.11 Manual Dosage
1) Click on Manual Dosage in the left-hand column, the below page will open.
2) From this page you can enter the desired product, dosage, machine and whether you want flush to create a manual
dosage.
3) Clicking on Start will begin the dosage.
3.12 Calibration
1) Click on Calibration in the left-hand column, the below page will open.
2) Type – choose whether you want to calibrate a product or water.
3) Product/channel – select the product you want to calibrate or the channel for water.
4) Destination machine – choose which machine you want to calibrate to at the point of inlet. Although currently you must
choose a washer extractor, a 3-way valve can be installed to make the physical calibration easier.
5) Click on validate parameters to obtain KF value and Flow rate. (CAUTION! ALWAYS press "Validate Parameters" before
pressing the calibration button, even if you want to re-calibrate after having cleaned the channel, without entered a real
volume.)
6) Place chemical drum on measuring scales and Tare to obtain zero. Initiate by holding the calibration button, use the amount
grams obtained on the measuring scales calculating exact dosage by using the specific gravity (SG) of the product.
(Note: Chlorine's SG is higher than most standard products.)
7) Insert real obtained value and save.
8) Repeat for all products
Manual Dosage Settings
Calibration Settings
17