Step 2:Open the upper cleaning seal cover,Carefully clean
the inner wall of the measuring glass tube with a brush(Pay
attention not to overflow the water from the cleaning port,
which will cause the water to seep into the circuit board of the
measuring tank) 。
Step 3 : Repeat the second step repeatedly until the colored
substance adhering to the inner wall of the glass tube is
removed。
Step 4:Cover the measuring box, clean the sealing cover, and
power on。

8. Matters needing attention

The first start-up of the system takes 10 minutes to warm up 。
Incoming line voltage must be within the allowable range,Wiring
of incoming terminal must be firm and reliable
It is strictly prohibited to use in places with explosion,
flammability, corrosion of metal, damage of insulation and steam。
Maintain the original assembly mode of water inflow and water
outflow. The internal measurement is precision glass tube. Do not
use too much force when changing the installation, which will
cause the glass tube to break.
This test method is a national standard platinum cobalt
colorimetry, which can not necessarily meet the needs of on-site
chroma. It is suggested that the darkest color solution on site can
be used as the standard solution for sample mixing according to
Operating instructions for SD Series Online Color Meter