Calibration Procedure
2.4. Move the jaw as below, then record the reading.
Current coil
2.5. If the reading from step 2.3. and 2.4. have the difference > 0.1, adjust VR52 and
repeat step 2.3. and 2.4. until both readings keep the difference ≤ 0.1
2.6. Remove the EUT from the current coil
Calibrate Offset
3.
(ensure no signal is supplied on EUT's jaw)
3.1. Press
button to select DC A mode
3.2. Adjust VR54 and VR53 until display shows 0.00
3.3. After the offset adjustment in step 3.2., the display value from environment should be
within +/- 0.05
Rev. 002
Procedure for Item Number:
Page 3 of 7