Preamp power set to +/‐ 5 V (R16 = open, R18 = 0 Ohm/short).
VR1 and additional components installed for HV control (installed at the factory).
VR3 and additional components installed for temperature control (installed at factory).
J5 installed and connected to PA preamp with flex cable.
J7 installed. Unlike the XR100 that has a separate signal out, the PA preamps send the
signal through the power flex cable. As a result, the signal must be taken from the PC5 J7
connector and then connected to the user supplied signal processor.
Input power is applied either to J1 with the AC adapter or to J4 from a power supply.