AmScope B200 Bedieningshandleiding - Pagina 13
Blader online of download pdf Bedieningshandleiding voor {categorie_naam} AmScope B200. AmScope B200 16 pagina's. B series biological microscope
Trouble
Switch on but bulb dark
Bulb is flickering or
brightness is unsteady
Brightness of view field
isn't enough or is
uneven
Image isn't clear
Causation
Plug is unreliable
Bulb is broken
Fuse is broken
Bulb is unstable
Bulb is broken
Bulb specification doesn't
meet the requirement
Brightness isn't adjusted
correctly
Objective isn't in correct
position
The size of iris aperture is too
small
Lens (objective, eyepiece,
condenser, light collector) has
dust
Position of condenser is too
low
Cover glass of specimen
doesn't meet the requirement
Cover glass of specimen isn't
in up direction
Remedy
Plug in again
Change bulb
Change fuse
Insert it again
Replacing bulb
Replacing bulb
Adjust rotation
potentiometer
Make the objective in
correct position
Adjust the size of iris
aperture
Clean it
Higher condenser
Use required thickness
cover glass (0.17mm)
Place specimen correctly
11