Drawell TD50002D Gebruikershandleiding - Pagina 5

Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Drawell TD50002D. Drawell TD50002D 8 pagina's. Precision balance

CALIBRATION
 The purpose for this operation is to calibrate the balance so as to achieve
the best performance in case there is obvious tolerance error in weighing
or the balance is located in different gravity due to different latitude.
 it is suggested to warm up the balance over half an hour before calibration.
 Remove all loads from the platter, press key TARE to clear the readings to
0.000g.
 Press and hold key CAL for 3 seconds, release the key CAL when "---
CAL---" is shown in the display, a figure of standard weight will flash in the
display, put a standard weight on the platter accordingly, standby state "---
---" will be shown in the display for a few seconds before the standard
weight figure is shown in the display, remove the weight, "------" standby
state will remain for a few seconds before the balance enter stable
weighing mode, showing 0.000g.
 It is recommended to make the operations twice to achieve the best
calibration result.
 Linearity Calibration, Press and hold key CAL for 3 seconds, release the
key CAL when "---CAL---" is shown in the display, after a short presence of
"- - - - ", a standard weight figure will flash in the display, press and hold
key PCS to enter linearity calibration mode, follow the instructions to
proceed linearity calibration.
WEIGHING
When the balance is warmed up and calibrated, 0.000g is shown in the display
indicating weighing mode, put the object on the platter, when stable, the weight
of the object is shown in the display.
Note, symbol O is stable indictor, when it stops flash, the balance is stable
TARING
When a loader is put over the platter, its weight is shown in the display, press
key TARE, 0.000g will be shown in the display, indicating the weight of the
loader is deducted, put the object into the loader, when stable, the figure
shown in the display is the weight of the object.

PIECE COUNTING

Remove all loads from the platter, press key TARE to clear the readings in the
display, press key PCS to enter counting mode, "- -COU- -" "- - - -" and figure
10(default sampling quantity) will in turns be shown in the display, keep
pressing key PCS to select sampling quantity from 10,25,50,100 up to
500,release key PCS when the desired quantity flashes in the display, put a
sample on the platter accordingly, press key PCS again to save setting, the
counting mode setting is finished, in counting mode, weighing unit will change