4.To memorise the new set point value push the SET key again or wait 10s.

8.1.4 How to start a manual defrost (only xr02cx)

Push the DEF
key for more than 2 seconds and a manual defrost will start

8.1.5 How to change parameter value

To change the parameter's value operate as follows:
1.Enter the Programming mode by pressing the SET+
blinking).
2.Select the required parameter. Press the "SET" key to display its value
3.Use
or
to change its value.
4.Press"SET"to store the new value and move to the following parameter.
To exit: Press SET+
NOTE: the set value is stored even when the procedure is exited by waiting the time-out to expire.

8.2 parameters

Hy Differential: (0,1°C ÷ 25°C / 1°F ÷ 45°F) Intervention differential for set point. Compressor Cut
IN is SET POINT + differential (Hy). Compressor Cut OUT is when the temperature reaches the
set point.
LS Minimum SET POINT: (-55°C÷SET/-67°F÷SET): Sets the minimum value for the set point..
US Maximum SET POINT: (SET÷99°C/ SET÷99°F). Set the maximum value for set point.
ot
First probe calibration: (-9.9÷9.9°C /-17÷17°F) allows to adjust possible offset of the first
probe.
od Outputs activation delay at start up: (0÷99min) This function is enabled at the initial start up of
the instrument and inhibits any output activation for the period of time set in the parameter.
AC Anti-short cycle delay: (0÷50 min) minimum interval between the compressor stop and the
following restart.
Cy Compressor ON time with faulty probe: (0÷99 min) time during which the compressor is active
in case of faulty thermostat probe. With Cy=0 compressor is always OFF.
Cn Compressor OFF time with faulty probe: (0÷99 min) time during which the compressor is OFF
in case of faulty thermostat probe. With Cn=0 compressor is always active.
CH Kind of Action: cL= cooling action; Ht = heating action.
Display
CF Measurement unit: (°C÷°F) °C =Celsius; °F =Fahrenheit. WARNING: When the measurement
unit is changed the SET point and the values
have to be checked and modified if necessary).
rE Resolution (only for °C):(dE ÷ in) dE= decimal between -9.9 and 9.9°C; in= integer
dy Display delay: (0÷15 min.) when the temperature increases, the display is updated of 1 °C/1°F
after this time.
or wait 15s without pressing a key.
keys for 3s ("°C" or "°F" LED starts
8