BD 9003TIG03-G Snelle referentiehandleiding - Pagina 12
Blader online of download pdf Snelle referentiehandleiding voor {categorie_naam} BD 9003TIG03-G. BD 9003TIG03-G 18 pagina's. Volumetric pump

Display
AIR IN LINE
Single air bubble exceeds alarm limit.
Set not fitted correctly into air in line
detector.
AIR IN LINE
Accumulated air bubbles exceeds
alarm limit.
DOOR OPEN
Door was opened during an infusion.
DOWNSTREAM OCCLUSION
A blockage has occurred downstream.
NO FLOW
Flow sensor detects no flow.
DO NOT USE
Internal error.
0000CF04424.indd 12
Action
• Ensure set is fitted correctly in the air-in-
line detector.
• Inspect infusion set for air bubbles and
take appropriate action.
• Remove air according to hospital policy.
• Check fluid level in container.
• Check fluid level in drip chamber.
• Restart infusion.
• Inspect infusion set for air bubbles and
take appropriate action.
• Remove air according to hospital policy.
• Check fluid level in container.
• Check fluid level in drip chamber.
• Restart infusion.
• Close door or clamp infusion set using
roller clamp.
• Restart infusion.
• Check fluid path between Pump and
patient for clamps, connectors, kinks or
blockages.
• Examine access site for signs of
complications e.g. redness, swelling, pain,
heat.
• Check flow sensor.
• Check fluid level in container.
• Ensure all clamps above Pump are open.
• Check fluid level in drip chamber.
• Ensure that the bag spike is inserted
correctly.
• Check flow sensor is clean.
• Remove Pump from use.
0000CF04424 Iss. 1 12/18