This guide contains instructions for installation of the BD CSampler™ Plus.
Installation Overview
The following figure shows steps for installation of the BD CSampler Plus.
Install
Validate
cytometer
cytometer
Before You Begin
Install and validate the BD Accuri™ C6 Plus flow cytometer.
1
Inspect the BD CSampler Plus accessory kit contents to ensure delivery of all
2
components. In addition to the BD CSampler Plus module, the box should also
contain the following.
• BD CSampler Plus module
• BD CSampler Plus mat
• Mounting bolts (3)
• Installation and removal tool
• 24-tube rack
• BD Accuri C6 Plus and BD CSampler Plus Software and Documentation CD
(not pictured)
Preparing the Cytometer
Shut down the cytometer.
1
WARNING
Failure to power off the cytometer before installing the BD CSampler Plus could damage the
BD Accuri C6 Plus or the BD CSampler Plus or both.
Loosen the thumb screw and remove the sample stage.
2
Class 1 Laser Product.
For Research Use Only. Not for use in diagnostic or therapeutic procedures.
© 2016 BD. BD, the BD Logo and all other trademarks are property of
Becton, Dickinson and Company.
23-18172-00
Plus Installation Guide
Inspect
Prepare
BD CSampler
cytometer
Plus
1
Install
Install
BD CSampler
software
Plus