Manual 24700GBM_ARC6
Then press <SELECT> and the display will move on to
received volume input. The first figure will "blink" to
show that it is changeable. Press <RESET> to change
the value press <SELECT> to change to the next figure.
In this example we received 1.6852 L so we move to
second figure with <SELECT> and change it from 5 to
6 with <RESET> move to third figure and change it
from 6 to 8 and so on.
When the last figure is blinking and <SELECT> is
pressed the display will change to show the calculated
calibration factor. If <RESET> is pressed the new
calibration is cancelled.
If <SELECT> is pressed the calibration is saved and
then the meter goes back to normal dispense mode. The
calibration is done.
If the calculated factor exceeds +/- 10% the error
message "E3" will be shown for a while and then it will
go back for a new calibration dispense.
Unit used for measuring
5.5.
The meter can be set to use either [L]itre, [G]allon, [P]int or [Q]uater. The unit used is shown after the value.
5.5.1.
How to set the unit
Press <RESET> and <SELECT> buttons simultaneous
and hold them down for 5 s. If successfully done the
display will change to this, showing that the meter is in
configuration mode. Now release the buttons.
Press <SELECT> button to get into "Set unit" mode. The
display will change to.
Scroll through the different units using the <RESET>
button. L, G, P, Q, L,...... . When the wanted unit is
shown release all buttons and wait. The will save and go
back to "ConF" after a short time.
To get back to normal operation mode wait a short time
or press <RESET> button. When the display looks like
this it is done.
◄ ▬ ►
◄ ▬ ►
▬ ►
▬ ►