3.8 Cooling

Atik cameras are thermally stabilized to allow your CCD to output the best result that
it can deliver. The very low readout noise, combined with efficient cooling, means that
dark frames are not necessary in most situations, allowing you to spend more time
imaging and less time taking calibration frames. If very long exposures are to be used
or high precision data is required (exo-planet searches, for example), dark frame
libraries will be easy to maintain.
The cooling delta varies between models and full specifications are available in the
Technical Information Summary on page 7. Please note: the cooling delta is how far
below the ambient temperature the camera can achieve, rather than a fixed
temperature it will cool to.
The Atik 314L+ has a regulated cooling system, meaning that the CCD is kept at the
desired temperature throughout the imaging session. After switching on the camera,
it is advisable to allow at least 2 minutes before taking images in order to allow
the temperature to stablilize.

3.9 Replacing the desiccant

Your camera includes a high-performance molecular
sieve desiccant tablet which is used to avoid
condensation in the CCD chamber. Although this
desiccant will last a long time it may eventually need
replacement or recharging, in which case please follow
the instructions below:
-
To replace the desiccant, open the desiccant port with the
supplied plastic tool and a screwdriver and take out the used tablet.
-
To recharge the desiccant, place the tablet in an electric oven
at 200ºC for 90 minutes. Take the tablet out the oven (CAUTION: tablet
will be very hot) and allow it to cool down.
-
Place the recharged tablet back in the camera, replace the
port cover and tighten (no need to over- tighten). Wait 24 hours before
connecting the camera again.
You will notice that there is a filter inside the desiccant chamber. The purpose of this
is to avoid contamination of the chamber when the desiccant is being replaced. The
filter is very fragile and should not be touched.
Page 7
Desiccant Port