Andesfit Health ADF-B06 Handmatig - Pagina 2

Blader online of download pdf Handmatig voor {categorie_naam} Andesfit Health ADF-B06. Andesfit Health ADF-B06 3 pagina's. Bluetooth 4.0 pulse oximeter

Alert Sound Setting
Hold the power-on key/functional key while the Oximeter is in powered-on state.
Parameter setting interface 1 is displayed, as shown in the following figure. Move
"*" to the corresponding option, and hold the functional key to set Alm to on and
set Beep to off. When Alm is set to on and the measured values of the blood
oxygen saturation and pulse rate go beyond the upper limit or lower limit, the
Oximeter gives off an alert sound. When Alm is set to off and the measured values
go beyond the limit, the Oximeter will not give any alert sound. When Beep is set to
on, a tick will be heard along with pulse beats during pulse rate measurement.
When Beep is set to off, no sound will be output along with pulse beats during
pulse rate measurement. While the "*" symbol stays on the Restore option, hold
the functional key to restore factory settings.
Brightness Setting
On parameter interface 1, press the functional key to select the Brightness option
and then hold the functional key to set the brightness to a value ranging from 1 to 5.
The greater the value, the greater the brightness of the screen.
Power saving function
The ADF-B06 has a power saving feature. It can save power through reducing
screen brightness automatically. During measurement the brightness will reduce to
level 1 automatically after 150 seconds without pressing the function key, to make
it recover to original brightness please press the function key.
Alert Range Setting
On parameter interface 2, press the functional key to switch between options. On
this interface, you can set the upper limit and lower limit of SpO2 Alm and PR Alm.
While the "*" symbol stays on the +/- option, hold the functional key to set the
option to + or -. In + mode, select the corresponding option and hold the functional
key to increment the upper or lower limit; in - mode, hold the functional key to
decrement the upper or lower limit. Move "*" to the Exit option, and hold the
functional key to return to the monitoring interface.
Operation Guide
Switch on the power supply of the pulse Oximeter
Stick one finger completely into the rubber pad cavity of the Oximeter, keep the
fingernail surface upward, and release the clip. Then press the power key to power
on the Oximeter.
If you do not yet completely insert
your finger into the cavity, the
measurement result may be
inaccurate.
Do not vibrate your finger during
measurement. Preferably, ensure
that your body does not move. After
the readings become stable, read
the measured values of
oxygen saturation and the
pulse rate on the screen.
The Oximeter will automatically shut down 10 seconds later after your finger leaves
away.
Connect device via Bluetooth 4.0
Prior to first use, download and install "Andesfit" application to your device.
(Bluetooth 4.0 capabilities) For details, see quick start guide for the phone / tablet.
About the Display
Continuously press the functional key in the monitoring process. The monitored
data and the display mode will be cyclically displayed on the OLED screen in two
different ways (large fonts and plethysmogram) and four directions, as shown in
the following figure.
Replace the batteries when the battery capacity is insufficient and the symbol
flickers on the screen.
Cleaning
Power off the instrument and remove the batteries before cleaning. Ensure that
the appearance of the instrument is neat, dust-free, and dirt-free. Clean the outer
surface of the instrument (including the OLED screen) using 75% medical alcohol
and a piece of dry soft cloth.
Caution: Avoid liquid flowing into the instrument during cleaning.
Caution: Do not immerse any part of the instrument into any liquid.
Disinfection
Before measurement with the instrument, wipe the rubber finger pad using a piece
of dry soft cloth dipped with 75% medical alcohol. Clean the finger to be measured
using the medical alcohol for disinfection purposes before and after use.
Maintenance
Remove the batteries from the battery slot and properly store them if you do
l
not plan to use the Oximeter for a long period of time.
Store the Oximeter between 14 and 122°F (–10 to +50°C) and at humidity
l
levels no greater than 93%.
Periodically check the Oximeter for damage.
l
l
Avoid using the Oximeter in an environment with inflammable gases or using it
in an environment where the temperature or humidity is excessively high or
low.
l
Check the accuracy of the oxygen saturation and pulse rate readings by using
an appropriate calibration apparatus.
Technical Specifications
1. Dimensions: 62.0 mm (Width) × 37.0 mm (Depth) × 32.0 mm (Height)
Weight: 42.5 g (including the height of the two AAA dry batteries)
2. Peak wavelength range of the light emitted from the probe: red light 660 nm ±
3; infrared light 905 nm ± 5.
3. Maximum optical output power of the probe: 1.2 mW for infrared light (905
nm).
4. Bluetooth module: 4.0 (single-mode)
5. Normal working condition
Working Temperature
5°C to 40°C
(41°F to 104°F)
Relative Humidity
15% to 80%, non-condensing
Atmospheric Pressure
70 kPa to 106 kPa
Rated Voltage
DC 3.0 V
6. Default values and conditions of alert
Parameter
Value
Upper limit: 100
Oxygen saturation
Lower limit: 94
Upper limit: 130
Pulse rate
Lower limit: 50
When the alert switch is on and the
actual measured value goes beyond the
Alert condition
preset alert parameter range, the
Oximeter gives an alert sound.
7. Technical parameters
Parameter
Value
Oxygen
35% to 100%
Display range
saturation
Pulse rate
25 bpm to 250 bpm
Oxygen
saturation
Resolution
Pulse rate
1 bpm
Oxygen
±2% (70% to 100%)
Measurement
saturation
No requirement (≤ 69%)
precision
Pulse rate
±2 bpm
Oxygen
Upper limit: 50% to 100%
saturation
Lower limit: 50% to 100%
Alert range
Upper limit: 25 bpm to 250 bpm
Pulse rate
Lower limit: 25 bpm to 250 bpm
Oxygen
± 1% of the preset value
saturation
Alert error
The greater of ±10% of the preset
Pulse rate
value and ±5 bpm
PI
Weak PI
Min. 0.2%
Safety Type
Anti-electric-shock type: internal power supply device
Anti-electric-shock degree: BF-type application part
Waterproof grade: IP22
Storage and Transportation
Packaged products should be stored in well-ventilated rooms without
corrosive gas and with an ambient temperature of –10 to +50°C, a relative
humidity lower than 93% (without condensation), and an atmospheric
1%