Atkins AquaTuff 353 Gebruikershandleiding - Pagina 5
Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Atkins AquaTuff 353. Atkins AquaTuff 353 6 pagina's. Digital thermocouple thermometer warranty and user guide
351 / 352 / 353 SPECIFICATIONS
TEMPERATURE RANGES:
35100-K / 35200-K / 35300-K:
100° to 999°F ( 73° to 537°C)
35132 / 35135 / 35140 / 35141:
35232 / 35235 / 35240 / 35241:
35332 / 35335 / 35340 / 35341:
100° to 500°F ( 73° to 260°C)
ACCURACY:
35100-K / 35200-K / 35300-K:
With factory calibration, ±0.5F° (±0.3C°) over entire
measurement range at ambient temperatures
between 68° to 86°F (20° to 30°C)
35132 / 35140 / 35141:
35232 / 35240 / 35241:
35332 / 35340 / 35341:
With factory calibration, ±0.9F° (±0.5C°) Total
System Accuracy over entire measurement range
at ambient temperatures between 68° to 86°F
(20° to 30°C)
35135 / 35235 / 35335:
Surface probe temperature errors for draft free, flat,
clean oiled surfaces with 2.2 lbs.(1 Kg) of pressure
are typically within +2.7°F and 6.3°F (+1.5°C and
3.5°C) without thermostat recycling.
RESOLUTION:
35100-K / 35300-K:
35132 / 35135 / 35140 / 35141:
35332 / 35335 / 35340 / 35341:
Tenth degree (.1)
35200-K / 35232 / 35235 / 35240 / 35241:
Tenth degree / whole degree (.1/1), user selectable
BATTERY LIFE:
1800 Hours (2) AAA Alkaline batteries
AMBIENT TEMPERATURE LIMITS:
32° TO 122°F (0° TO 50°C)
TRACEABILITY:
We certify that this instrument has been calibrated
to temperature and/or electrical standards traceable
to the U.S. Department of Commerce National
Institute of Standards and Technology.
CALIBRATION:
Under normal operating conditions, the unit should
not require re calibration.
8
351 / 352 / 353 TROUBLESHOOTING
TROUBLESHOOTING
Our Customer Service Department is available
Monday through Friday 8:00 AM to 5:00 PM EST
at 800-835-5011.
1) Unit Does Not Turn On:
Batteries are installed incorrectly or dead.
Check batteries are installed according to picture on
rear label. Replace batteries.
2) Display Shows "Prob": Probe is broken (Open
thermocouple see "Test for Broken Probe" below).
3) Test for Broken Probe: Pull the probe connector out
partially to expose the blade contacts. Place a metal
paper clip across connector blades to short contacts. If
instrument displays temperature, the probe is broken.
Replace Probe.
4) Battery Icon Appears on Display: Low battery.
Replace batteries.
5) Display Shows "bAtt" and Instrument Shuts Off:
Replace batteries.
6) Display Shows "LO" or "HI": Temperature measured
is outside measurement range of instrument.
351 / 352 / 353 WARRANTY
WARRANTY: Cooper Atkins thermometers are backed
by a 5 year Instrument Warranty and 1 year Probe
Warranty on defects in workmanship and materials.
9