Ankom A2000 Service Procedure
Note: If your A2000/A2000I does not have a SICK Module,
see Service Procedure 22.
Replacing the Glass Sensor Tip:
1.
Using a screw driver, remove the clip holding the fiber
optic cable in the sensor. Be careful not to lose the
clip.
2.
Unplug the fiber optic cable.
3.
Follow the cable up to the plastic fitting in the vessel.
4.
Gently unscrew the plastic fitting. (Allow the cable to
turn freely.)
ANKOM Technology 2052 O'Neil Rd Macedon, NY 14502 PH: (315) 986-8090 FAX: (315) 986-8091
Level Sensor (SICK) Component Replacement
A2000/A2000I
5.
Replace the fitting/cable with the new fitting/
cable. Be careful not to damage the glass tip
protruding from the fitting.
6.
Tighten the fitting firmly by hand in the
vessel, then tighten the fitting 1/4 turn with a
wrench which will compress the O-ring to
make the seal.
Reminder:
The new glass sensor has to be used with the new
fiber optic sending unit. Do not mix and match the
two components.
7.
See Test the Level Sensor Section in Service
Procedure #6.
011
Service Procedure
(Page 1 of 3) 08/05/2014