Note: Follow steps 1-8 for level sensor module replacement and 8-12 for Fiber optic cable replacement. (1-12 for
both)
1. Unplug the Instrument.
2. Unscrew the four screws located on each side of the cabinet back. Lift the cabinet back while removing it from
the instrument.
3. Open the electrical box by removing the two screws located on top of the instrument.
4. Loosen the two retainer screws using a flat head screwdriver (figure 1)
5. Gently pull fiber optic cable from level sensor.
ANKOM Technology • 2052 O'Neil Road • Email: [email protected]• Phone: 315-986-8090 • Fax: 315-986-8091
Replacing X59 Analog level sensor
and Fiber optic cable with glass tip
XT15/XT15I
Service Procedure
JAD 4/2/2019
Figure 1
Figure 2
X5S012
pg. 1 of 4