Druck InstruMetrics ADTS542F Veiligheids- en installatiehandleiding - Pagina 12
Blader online of download pdf Veiligheids- en installatiehandleiding voor {categorie_naam} Druck InstruMetrics ADTS542F. Druck InstruMetrics ADTS542F 16 pagina's. Air data test set
1.
Clean the external surfaces of the unit using lint-free cloth and a soft brush. Remove
stubborn dirt using a damp cloth and mild detergent and allow to dry.
2.
Clean the electrical connector using a soft brush. Remove stubborn dirt using a damp cloth
and mild detergent and allow to dry.
4. Specification
4.1 Operating Environment
WARNING
atmospheres.
Storage temperatures
Operating
temperatures/environment
ADTS Touch universal power
supply
Power supply
Electrical safety
Operating altitude
Ingress protection
Copyright 2014 Baker Hughes Company.
10 | ADTS Safety and Installation Guide–English
This equipment is not rated for use in potentially explosive
ADTS542F:
ADTS552F )
ADTS553F ) -20°C to 70°C (-4° to 158°F)
ADTS554F )
ADTS542F:
ADTS552F )
ADTS553F )
ADTS554F )
Operating humidity: 5% to 95% RH (condensing)
Indoor & Outdoor (non-continuously) use
Operating temperature: 10°C to 50°C (50° to 122°F)
Operating humidity: 5% to 95% RH (non-condensing)
Universal input Mains Power Supply.
Class 1 equipment requiring a protective earth connection.
TXAH 250V HRC Time-delay (T) 250Va.c. rated fuses (X = amp
rating of fuse).
Overvoltage category II, Pollution degree 2
The supplied mains/PSU cable protective earth/ground conductor
must be connected to the power supply protective earth/ground
system.
EN 61010-1, UL 61010-1, CSA 22.2, No. 61010-1 and IEC 61010-1
Maximum 7500ft (2300 metres)
ADTS542/552/553/554 IP33, ADTS Touch IP65 (EN60529)
-20°C to 70°C (-4° to 158°F)
-5°C to 50°C (32° to 122°F)
-5°C to 50°C (32° to 122°F)