Druck IO620-CHARGER Gebruiksaanwijzing - Pagina 2
Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} Druck IO620-CHARGER. Druck IO620-CHARGER 2 pagina's.
Preparation
To use this charger, make sure you have
these items:
•
IO620-PSU: 5 V DC power supply. A
standard item with the DPI 620G
calibrator.
Alternative: A USB mini Type B cable
(IO620-USB-PC); not a standard item.
•
IO620-BATTERY: Li-ion battery.
Charge Method
Charge Time (to full
capacity)
5 V DC power supply
6.5 hours
USB mini Type B
13 hours (with a
cable
500 mA supply)
K0455 Revision A
4
Assembly and Operation
In a dry location, put the charger on a flat
surface near an applicable power socket: AC
wall socket or USB computer socket. Then:
1. Connect the power supply to the charger:
(a) 5 V DC power supply OR
(b) USB mini Type B
2. Put the battery in the charger until the
guides (c), (d) and (e) hold it in position.
LED Indications
Condition
Flashing red and green
No battery
LEDs
Continuous red LED
Charging
Continuous green LED
Charging
complete
5
Red LED
1
Green
LED
a
OR
b
2
c
d
c
6
Specification
Charge Temperature: 0 to 40°C (32 to
104°F). When the charger senses the
temperature is outside this range, it stops
charging.
Storage Temperature: -20 to 70°C (-4 to
158°F).
Humidity: 0 to 90% relative humidity (RH)
non-condensing.
EMC: BS EN 61326-1:2013
Electrical Safety: BS EN 61010:2010
Size (L:W:H); weight: 97 x 83 x 22 mm
(3.8 x 3.3 x 0.7 in); 50 g (1.8 oz)
Customer Service
Visit our web site: Druck.com
e
Copyright 2008 Baker Hughes Company. This material contains one or more
registered trademarks of Baker Hughes Company and its subsidiaries in one or
more countries. All third-party product and company names are trademarks of their