4. Making Opto-electronic Component Measurements
3. The response is the absolute detected power and will have a similar shape to the |S21| plots in the
previous section but the numerical levels will be different.
Figure 3.
Example Plot of the Functional Test at +5 and +2 dBm Optical Power Levels
An example plot of the functional test is shown in
power (blue curve) and +2 dBm optical power (for reference, red curve). The -0271/0272 and -0240 systems will
have values somewhat lower than the red curve (2-5 dB for the -0271/0272 and 5-10 dB lower for the -0240).
4.
Making Opto-electronic Component Measurements
The typical measurement scenario with the ME7848A opto-electronic network analyzer system is to analyze
the conversion behavior of E/O or O/E devices or to study the RF envelope behavior for signals passing through
O/O components. All of these measurements are done from an RF perspective by the VNA with the optical
carrier (provided by the laser) acting as a parameter. Much more measurement information is available in the
VectorStar Calibration and Measurement Guide, 10410-00318, (with an entire chapter devoted to this
particular measurement class) but some salient points will be summarized here.
• The VNA calibration establishes reference planes (for insertion loss and phase, group delay, and
reflection) at the ends of the RF cables (or wherever the calibration was performed). Thus a
measurement of, for example insertion loss, the structure in
conversion loss of the E/O and O/E modules and the optical interconnect.