Absolute Process Instruments Cecomp DPG2000BBL D4-M0 Series Instructies

Blader online of download pdf Instructies voor {categorie_naam} Absolute Process Instruments Cecomp DPG2000BBL D4-M0 Series. Absolute Process Instruments Cecomp DPG2000BBL D4-M0 Series 4 pagina's. Intrinsically safe digital pressure gauges, with min/max options

Absolute Process Instruments Cecomp DPG2000BBL D4-M0 Series Instructies
Cecomp
Intrinsically Safe Digital Pressure Gauges, with Min/Max Options
®
Agency Approval
Factory Mutual Approved Intrinsically Safe for Hazardous Locations
USA & Canada
IS Class I, Division 1, Groups A, B, C, D
T3C Ta = –40°C to 82°C; T4 Ta = –40°C to 66°C
CL I Zone 0 AEx/Ex ia IIC
T3 Ta = –40°C to 82°C; T4 Ta = –40°C to 66°C
Ranges and Resolution
See table below. Select range and default engineering unit.
Units may be changed to any listed under the same sensor range.
Resolution is fixed and limited to available display digits.
Display
3 readings per second nominal display update rate
4 digit LCD, 0.5" H and 5 character 0.25" H alphanumeric
BL:
Red LED backlight. Keypress activates backlighting for 1
minute if ambient light is insufficient.
Accuracy
Accuracy includes linearity, hysteresis, repeatability
Standard accuracy: ±0.25% of full scale ±1 least significant digit
HA accuracy option: ±0.1% FS ±1 LSD, see ranges for availability
Sensor hysteresis: ±0.015% FS, included in accuracy
Sensor repeatability: ±0.01% FS, included in accuracy
Auto Shutoff
5, 10 , 30 minutes, or on/off. User selectable 1 minute to 8 hrs
Memory Options
Min/max can be user configured to be individually enabled or dis-
abled, readings saved or cleared at power off
Std:
Min/max enabled
M0:
Min/max turned off in user setup
M1:
Peak reading only gauge for tire pressure
Calibration
Non-interactive zero, span, and linearity, ±10% of range Internal
setup/calibration buttons, internal lockout switch.
Sensor Ranges and Engineering Units
3 psig *
Res
15 psig vac *
3PSIG .001
100KPAVAC
6INHGG .001
0.1MPAVAC .0001
85INH2OG
1BARVAC .001
.1
50ZING
.01
1KGCMVAC .001
210GCMG
.1
1ATMVAC .001
150MMHGG
.1
15 psig
150TORRG
15PSIG
.1
200MBARG
.1
30INHGG
200CMH2OG
.1
400INH2OG
2000MMH2OG
1
240ZING
7FTH2O .001
1000GCMG
20KPAG
.01
760MMHGG
5 psig *
Res
760TORRG
5PSIG .001
1000MBARG
10INHGG
1000CMH2OG
.01
140INH2OG
.1
35FTH2O
80ZING
.1
100KPAG
350GCMG
.1
0.1MPAG .0001
260MMHGG
1BARG .001
.1
260TORRG
.1
1KGCMG .001
350MBARG
.1
1ATMG .001
350CMH2OG
±15 psig *
.1
3500MMH2OG
1
±15PSIVAC
12FTH2O
.01
15PSICPD
35KPAG
.01
±30INHGVAC
15 psia
Res
±400INH2OVAC
15PSIA
.01
±240ZINVAC
30INHGA
.01
±1000GCMVAC
400INH2OA
.1
±760MMHGVAC
240ZINA
±760TORRVAC
.1
1000GCMA
1
±1000MBARVAC
760MMHGA
.1
±1000CMH2OVAC
760TORRA
.1
±100KPAVAC
1000MBARA
±0.1MPAVAC .0001
1
1000CMH2OA
1
±1BARVAC .001
100KPAA
.1
±1KGCMVAC .001
0.1MPAA .0001
±1ATMVAC .001
1BARA .001
30 psia
1KGCMA .001
30PSIA
1ATMA .001
60INHGA
15 psig vac *
Res
850INH2OA
15PSIVAC
480ZINA
.01
30INHGVAC
.01
2100GCMA
400INH2OVAC
.1
1600MMHGA
240ZINVAC
.1
1600TORRA
1000GCMVAC
2000MBARA
1
760MMHGVAC
.1
2100CMH2OA
760TORRVAC
.1
200KPAA
1000MBARVAC
0.2MPAA .0001
1
1000CMH2OVAC
1
2BARA .001
C E CO MP
Div. of
Controls and Functions
Front button turns gauge on or off, zeros gauge reference gauges,
and cycles through min/max functions (if enabled)
Internal buttons for engineering unit selection, auto shutoff time,
min/max setup, calibration
Internal lockout switch to disable setup and calibration
Batteries
Two 1.5 V AAA Panasonic LR03 alkaline cells
B:
Approx. 1000 hours
BL:
Approx. 150 to 1000 hours depending on backlight usage
Low battery symbol on display when batteries need replacement
Weight
9 ounces (approx.), shipping wt. 1 pound (approx.)
Housing Materials and Circuit Board Protection
Epoxy powder coated aluminum case, rear cover, and bezel. Front
and rear rubber gaskets, polycarbonate label.
Stainless steel stiffener plate to reinforce sensor area.
Conformal coating on circuit boards for moisture resistance.
Connection and Material
1/4" NPT male fitting. All wetted parts are 316L stainless steel.
Overpressure, Burst, Vacuum
2 X pressure range for 3 psi to 2000 psi sensors
5000 psig for ranges using 3000 psig sensor
7500 psig for ranges using 5000 psig sensor
Over-range display 112.5% FS: I – – – or I –.–.–.–
Under-range display (non-vacuum sensors): –Err
Vacuum service:
Burst: 4 X sensor pressure rating or 10,000 psi, whichever is less
Environmental
Storage temperature:
Operating temperature:
Compensated temperature: 32 to 158°F (0 to 70°C)
* -HA option not available
Res
30 psia
Res –15V100psig * Res
.1
2KGCMA .001
2ATMA .001
30 psig
Res
200INHGVAC
30PSIG
.01
2770INH2OVAC
60INHGG
.01
1600ZINVAC
Res
850INH2OG
1
5200MMHGVAC
480ZING
5200TORRVAC
.01
.1
.01
2100GCMG
1
700KPAVAC
.1
1600MMHGG
1
0.7MPAVAC .001
.1
1600TORRG
1
2000MBARG
1
1
.1
2100CMH2OG
1
.1
70FTH2O
.01
100 psig
1
200KPAG
.1
0.2MPAG .0001
1
.01
2BARG .001
2770INH2OG
.1
2KGCMG .001
2ATMG .001
60 psig
Res
5200MMHGG
60PSIG
.01
5200TORRG
120INHGG
.1
7000MBARG
Res
1660INH2OG
7000CMH2OG
1
.01
960ZING
1
.01
4200GCMG
1
.01
3100MMHGG
1
3100TORRG
1
1
.1
4100MBARG
1
1
4200CMH2OG
1
1
140FTH2O
.1
–15V200 psig * Res
400KPAG
1
.1
1
0.4MPAG .0001
1
4BARG .001
400INHGVAC
.1
4KGCMG .001
5500INH2OVAC
4ATMG .001
3200ZINVAC
100 psia
Res
1400KPAVAC
100PSIA
.1
1.4MPAVAC .001
200INHGA
.1
Res
2770INH2OA
14KGCMVAC
1
.01
1600ZINA
1
.01
7000GCMA
1
200 psig
1
5200MMHGA
1
5200TORRA
.1
1
1
7000MBARA
1
5500INH2OG
1
7000CMH2OA
1
1
700KPAA
.1
0.7MPAA .0001
1
1
7BARA .001
.1
7KGCMA .001
7ATMA .001
A
P
BSOLUTE
ROCESS
15 psia, ±15 psig, 15 psig, 30 psia,
100 psig, 100 psia, 200 psig sensors
–40 to 203°F (–40 to 95°C)
–4 to 185°F (–20 to 85°C)
Range codes are rounded off
300 psig
Res
100PSIVAC
.1
300PSIG
.1
100PSICPD
.1
610INHGG
.1
4800ZING
.1
1
700FTH2O
.1
1
2000KPAG
1
2MPAG .001
20BARG
1
.01
1
20KGCMG
.01
20ATMG
.01
7BARVAC
.01
500 psig
Res
7KGCMVAC
500PSIG
.01
.1
7ATMVAC
.01
1020INHGG
Res
1150FTH2O
100PSIG
.1
3500KPAG
200INHGG
3.5MPAG .001
.1
1
35BARG
.01
1600ZING
1
35KGCMG
.01
7000GCMG
1
35ATMG
.01
1000 psig
Res
1
1
1000PSIG
1
2040INHGG
2300FTH2O
1
230FTH2O
.1
7000KPAG
700KPAG
.1
7MPAG .001
0.7MPAG .0001
70BARG
.01
7BARG .001
70KGCMG
.01
7KGCMG .001
70ATMG
.01
7ATMG .001
2000 psig
Res
2000PSIG
200PSIVAC
4070INHGG
.1
200PSICPD
.1
4600FTH2O
.1
14MPAG
.01
1
140BARG
.1
140KGCMG
1
.1
1
140ATMG
.1
3000 psig
Res
14BARVAC
.01
3000PSIG
6100INHGG
.01
14ATMVAC
.01
6900FTH2O
Res
20MPAG
.01
200PSIG
.1
200BARG
.1
400INHGG
200KGCMG
.1
.1
1
200ATMG
.1
3200ZING
1
5000 psig
Res
480FTH2O
.1
5000PSIG
1400KPAG
35MPAG
1
.01
1.4MPAG .001
350BARG
.1
14BARG
.01
350KGCMG
.1
14KGCMG
340ATMG
.01
.1
14ATMG
.01
1220 American Way Libertyville, IL 60048
I
NSTRUMENTS
DPG2000B D4 and M0, M1 Series
O ±0.25% Test Gauge Accuracy
O 316L Stainless Steel Sensor
O All Metal Housing
How to Specify
DPG2000B range -D4-time-options
DPG2000BBL range -D4-time options
DPG2000B range -D4-M0-time-options
DPG2000BBL range -D4-M0-time-options
DPG2000B range -D4-M1-time-options
DPG2000BBL range -D4-M1-time-options
1
Range—See table at left. Select a range code for default units.
Please specify if vacuum gauge requires a minus sign.
1
psi = PSI
torr = TORR
inHg = INHG
mmH
O = MMH2O
2
oz/in
2
= ZIN
kg/cm
2
= KGCM
inH
O = INH2O
g/cm
2
= GCM
2
ftH
O = FTH2O
kPa = KPA
2
mmHg = MMHG
MPa = MPA
1
1
G = gauge reference pressure
1
CPD = compound; inHg vacuum, psi pressure
VAC = gauge reference vacuum
A = absolute reference
Time—auto shutoff time
5
5 minutes. Default if not specified.
1
10
10 minutes
1
30
30 minutes
1
ON
No auto shutoff. On/off via front button.
1
xH
Custom shutoff time where x = up to 8 hours
Options—add to end of model number. See price list for details.
HA
High accuracy, ±0.1% FS ±1 LSD. See range table.
PM
Panel mount, 4.1" x 4.1"
TP
Top port, gauge port on top of case
1
CD
Calibration data; 5 test points and date
1
NC
1
NIST traceability documentation, 5 points and date
TP
Top gauge port. Primarily used with tire pressure
applications. Not available with NEMA 4X models.
1
Accessories—order separately
1
RB
1
High visibility orange rubber boot protects gauge
for portable applications. Not available with NEMA
4X models.
SCR14SS
Filter screen fitting keeps debris out of gauge
sensor. For food vacuum packaging applications.
1
303SS body, 100 micron 304SS screen.
CON14SS
Quick connector to install or remove gauge with-
out tools. 304 stainless steel, urethane seal.
cecomp.com
800-942-0315
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
Made in USA
Quick Link
cecomp.com/is
Display Memory
-
Min/max
backlit
Min/max
-
Min/max off
backlit
Min/max off
-
Peak read
backlit
Peak read
mbar = MBAR
bar = BAR
cmH
O = CMH2O
2
atm = ATM
© 08-18