4.Operation Method

4.1 Turn on and off

Adapter: Insert the plug of AC Adapter into the power supply hub which is on the left of
the Display Indicator, and insert the AC Adapter into AC socket.
Battery: Open the battery cover which is on the right of the Display Indicator, load 4 x AAA
alkaline cells into the cell box (note the polarity) and then close the battery cover.
Empty the platform, press
Maximum Capacity and Division. After the screen displays "0", you can begin to weigh the
object.
When power-on, press

4.2 Connecting the Load Cell

The connection of the four wire load cell and the indicator:
Connect +EXC with the red wire of the indicator's cable.
Connect –EXC with the black wire of the indicator's cable.
Connect +SIG with the green wire of the indicator's cable.
Connect –SIG with the white wire of the indicator's cable.
The indicator provides driving voltage through +EXC wire and –EXC wire. +SIG wire and
–SIG wire will transmit the change of signal voltage to the indicator.
*To avoid short circuit, please make sure the welding parts of all the wires covered by
the heat shrinking tube.
Figure: Connection of the Load Cell and the indicator
2 |
A n y l o a d D D - K C 1 M a n u a l V ( 1 6 0 3 )
button once, it will show all the characters then the