Duomo CellAir Gebruikershandleiding
Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Duomo CellAir. Duomo CellAir 5 pagina's. Co2 detector
Application
Thankyou for selecting the Duomo CellAir
monitor. It is designed to detect the presence of
carbon dioxide in the ambient air to protect people
in confined spaces. High concentrations of CO
confined spaces are dangerous, and may lead to
health problems ranging from headaches and
fatigue to asphyxiation and death. CellAir
audible alarm and visual indication which will
activate when CO
concentration reaches a pre-set
2
level. Detection of high levels of CO
activate a relay that could be used for a fan to
ventilate the confined space and reduce CO
concentration in the area. CellAir
used in CO
storage areas, breweries, wineries,
2
cellars, beverage dispensing areas and fast food
outlets.
Package Contents
SEU (Sensor Unit), RDU (Remote Display Unit), panel
holders, network cable connector, communication
cable (8m length), fixtures and fittings, user manual.
w: www.duomo.co.uk
-
TM
has an
TM
will also
2
can be widely
TM
-
CellAir
NDIR (Non-Dispersive Infrared) sensor
Measures up to 50,000ppm
Can handle up to 3 remote display units;
1 up to 25m away, 3 up to 8m apart
Large LCD display indicates ambient
CO
concentration and temperature
2
Relay output to automatically control
a fan to ventilate confined spaces
Audible and visual alarm indications
IP54 water proof protection
Specifications
CO
Specification
CO
2
2
Measurement range - 0 - 50,000ppm
Accuracy - ±100ppm or ±5% of reading
Temperature Specification
in
2
Measurement range - 0
Alarm Levels
Alarm 1 - 0.5 / 1 / 1.5 / 2%
Alarm 2 - 1.5 / 2 / 2.5 / 3 / 3.5 / 4%
Operating Conditions
Operating temperature - 0
Humidity range - 0 ~ 95%RH
Storage Conditions
2
Storage temperature - -20
Power Supply & Relay Output
AC adapter - 110/220 VAC
AC input - 100 ~ 240 VAC, 50/60Hz
AC output - 6 VDC, 1.8 W
Relay output - 30 VDC or 250 VAC, max. 2A, SPST, NO
Accuracy
Annual drift < 20ppm / year @ 400ppm
Calibration
User calibration via onboard menu system
Recommended calibration interval - 2 years
t: 01905 797989
-
CO
Detector
TM
2
o
o
C - 50
C (32
o
o
C - 40
C (32
o
o
C - 60
C (-4
f: 01905 774296
© Duomo (UK) Ltd. 2016
|
o
o
F - 122
F)
o
o
F - 104
F)
o
o
F - 140
F)