4B ADBT2 Bedieningshandleiding - Pagina 12
Blader online of download pdf Bedieningshandleiding voor {categorie_naam} 4B ADBT2. 4B ADBT2 16 pagina's. Adjustable depth bearing sensor temperature tester
FAULT
LED Display not Functioning
Testing Temperature Set
Buttons not Working
Battery Low Light Indicator /
Heater Block not Warming
ADB Sensor not Responding
as Expected to Testing
TROUBLESHOOTING GUIDE
Replace 9 Volt batteries.
Turn ADBT off and try again, if buttons are still not functioning
replace 9 volt batteries.
Replace AA batteries.
1. Make sure the ADB sensor probe is securely inserted all the way
into the tester's heater block.
2. Check alarm and shut down settings on control unit.
3. Replace ADB sensor and re-test. If sensor is still not responding
as expected, contact 4B.
PAGE 12
REMEDY