Aplisens PCE-28 Gebruikershandleiding - Pagina 14
Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Aplisens PCE-28. Aplisens PCE-28 20 pagina's. Pressure transmitters
IO.PCE-28....PC-29.Exd
PCE-28.Smart, PRE-28.Smart connection
„FL(...)" connection
Red wire
Black wire
PCE-28.Modbus, PRE-28.Modbus connection
„FL(...)" connection
Red wire
Black wire
Blue wire
Yellow wire (VB)
PC-29A, PR-29A, PC-29B, PR-29B connection.
„FL(...)" connection
Red wire
Black wire
Blue wire
9.2.
How to connect transmitters in potentially explosive areas
General principles in relation to connecting and using Exd transmitters should be
consistent with the rules and standards concerning equipment with flame-proof housing as
in point 7.1, including:
EN60079-14 - Electrical equipment in potentially explosive atmospheres. Part 14:
Electrical installation in risk areas (other than mines).
EN60079-17 - Electrical equipment in potentially explosive atmospheres. Part 17:
Inspection and operation of electrical installation in hazardous areas (other than mines).
Cables should be protected against damage by laying them in cable trays, lining
pipes, cable ladders, using durable fastenings etc.
Transmitters and equipment in a transmitter measurement loop should be
connected in accordance with explosion-proof standards and conditions of use in
hazardous areas.
Failure to comply with explosion-proof may result in an explosion and risk to
humans associated with it.
Transmitters should be powered by power supplies or other equipment that
ensures that there is reinforced insulation between primary and secondary
windings and which do not exceed 250V AC. It is the user who is responsible for
the provision of power supply in compliance with the aforementioned
requirements.
12
(+)
(-)
(+)
(-)
(VA)
(+)
(-)
(OUT)
„SGM(...)" connection
Red wire (+)
Black wire
(-)
Green wire (cable shield)
„SGM(...)" connection
Red wire
(+)
Black wire
(-)
Blue wire
(VA)
Yellow wire (VB)
Green wire (cable shield)
„SGM(...)" connection
Red wire
(+)
Black wire
(-)
Blue wire
(OUT)
Green wire (cable shield)
04.A.004