Apogee Instruments SQ-620 Gebruikershandleiding - Pagina 10

Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Apogee Instruments SQ-620. Apogee Instruments SQ-620 15 pagina's. Extended range pfd sensor
Ook voor Apogee Instruments SQ-620: Gebruikershandleiding (15 pagina's), Gebruikershandleiding (15 pagina's)

Apogee Instruments SQ-620 Gebruikershandleiding

OPERATION AND MEASUREMENT

Connect the sensor to a measurement device (meter, datalogger, controller) capable of measuring and displaying a
4-20 mA signal. To maximize measurement resolution and signal-to-noise ratio, the input range of the
measurement device should closely match the output range of the quantum sensor. DO NOT connect the sensor
to a power source greater than 24 V DC.
Wiring for SQ-624
Black: Ground (for sensor signal and input power)
Red: Input power (7-24 V DC)
White: 4-20 mA signal output
Clear: Shield/Ground
10