Apogee Instruments SQ-644-SS Gebruikershandleiding - Pagina 10
Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Apogee Instruments SQ-644-SS. Apogee Instruments SQ-644-SS 16 pagina's. Quantum light pollution sensor
Ook voor Apogee Instruments SQ-644-SS: Gebruikershandleiding (16 pagina's)
OPERATION AND MEASUREMENT
Connect the sensor to a measurement device (meter, datalogger, controller) capable of measuring and displaying a
4-20 mA signal. To maximize measurement resolution and signal-to-noise ratio, the input range of the
measurement device should closely match the output range of the quantum sensor. DO NOT connect the sensor
to a power source greater than 24 V DC.
Wiring for SQ-644
White: 4-20 mA signal output
Red: Input power (7-24 V DC)
Black: Ground (for sensor signal and input power)
Clear: Shield/Ground