CMP 6050 RETROFIT INSTALLATION INSTRUCTIONS
10.
F
Locate the four pre-drilled bezel assembly mounting holes on new control panel assembly (Figure
7). Drill through control panel door using a 3/16" drill bit and the bezel mounting holes as a guide.
11.
F
Mount bezel assembly to front of control panel door, using #10/32 x 1/2" screws included in the
Retrofit Hardware Package. Note: The display, buzzer, and stylus holder will be pre-installed at the
factory.
12.
F
Release bundled buzzer wires from the rear of the bezel assembly and connect wire labelled X4 to
terminal block X4, and wire labelled AL to terminal block AL on the new control panel assembly. If
wires are not attached to the rear of the buzzer, connect wire labelled AL to buzzer terminal La, and
wire labelled X4 to buzzer terminal N.
13.
F
Release upper wire bundle connected to new control panel assembly (Figure 9). Using wire labels
and terminal connectors provided, make connections for power and communications to the rear of
the display (Figure 10) as shown on schematics 231106 and 257349 page one.
Upper wire bundle
4
May 2013 |
257775R00
Figure 8
Figure 9
Lower wire bundle
CONVIRON