Cleaning

Disconnect the mains plug before cleaning the scale. Use only a damp cloth or an ordinary
disinfectant for cleaning. Do not use aggressive liquid cleaning agents, abrasive or acidic
detergents or white spirit.
Make sure that no liquid cleaning agent or water penetrates the scale and always follow the
manufacturer's instructions for use.

Faults and error messages

Below you will find a list of possible errors and faults. If necessary, these can be eliminated
with simple means. If the fault or error cannot be eliminated or other error messages occur,
please contact your service partner.
Faults
Fault
The scale has switched off
automatically.
The display
shows nothing.
The scale is not connected to the
mains.
Draughts or air movements
Weight display is
Vibrations
unstable.
Electromagnetic fields / static
charging
The scale is not levelled
properly.
The weighing
The scale is not operating in the
result is
specified temperature range.
obviously wrong.
Electromagnetic fields / static
charging
Zeroing is not possible if the
Zeroing is not
displayed weight exceeds +/- 2%
possible.
of the capacity of the scale.
LW300_LW300T-200520-Rev005-UM-en
key or cancel the procedure by pressing the

10. Care and maintenance

Cause
Measure
Switch on the scale
Use the optional mains adapter to
connect the scale to the mains.
Close the window.
Choose a different location.
Avoid vibrations.
Choose a different location.
Choose a different location. If
possible, switch off the interfering
device.
Adjust the support feet. Use the
spirit level to check the horizontal
position.
Operate the scale only in the
permissible temperature ranges.
Choose a different location. If
possible, switch off the interfering
device.
In this case, switch the scale off and
on again.
13