ADE M500660 Bedieningshandleiding - Pagina 10

Blader online of download pdf Bedieningshandleiding voor {categorie_naam} ADE M500660. ADE M500660 16 pagina's. Electronic wheelchair and multipurpose scales

ADE M500660 Bedieningshandleiding

Evaluation of the Body Mass Index

Compare the determined value which corresponds to those being used by the WHO.
BMI below 18,5
The person is under the regular standard and does not got enough weight. An increase of weight is
recommended. Please consult a professional.
BMI between 18,5 and 24,9
The person got a normal weight. If he feels well there is no need for a change.
BMI between 25 and 30
The person is slightly overweighed to pre-obese. A reduction of weight is indicated or, is necessary if
the state of health is already affected (e.g. diabetes, high end blood pressure, heart diseases).
BMI over 30
Reduction of weight is imperative. To visit a doctor is highly recommended.

Connecting additional devices

Connection of printer, PC or remote-display will be done with the RS-232 interface. For connection of addi-
tional devices kindly ask your local representative or authorized ADE dealer for further instructions.

Print out a record

The scales must be at standstill. The standstill status symbol
TER key to send the data to the serial interface.
Note: Printing is not possible if the net weight is 0. The print function only produces one print-out. A reset is
required to print out another copy.

Error Messages

Error Codes
Load cell wires to indicator are incorrectly
connected; or shorted, or opened; or
ADC/ load cell are being damaged
Weight reading exceeds the zero limit
after switching the scale on.
Weight reading is below the zero limit
after switching the scale on.
Weight reading exceeds load limit.
Weight reading below load limit.
Configuration or Calibration parameters
are not correctly set.
Capacity parameters are not correct.
Calibration error. The incoming signal of
loaded weight is too small, too big, unsta-
ble or un-linear.
Power cord is not properly connected.
Not turn on.
Power outlet not supplying electricity.
Batterie is discharged. Other failure.
Cannot zero the
Load on scale exceeds allowable limits.
display or will not
Load on scale is not stable.
zero when turned
Load cell damage.
on.
10
Description
must appear at the screen. Press the EN-
Solution
Make sure wires are in good order
and condition and correctly con-
nected. Replace load cell or ADC
chip. Service is required.
Make sure scale platform is empty.
Install platform on scale. Otherwise
the zero calibration/ adjustment
have to be done by authorized per-
sonnel.
Reduce load on scale until weight
value can be displayed.
Service by authorized personnel
required.
Contact your local dealer / supplier
for calibration purpose.
Contact your local dealer / supplier
for a new set-up.
Contact your local dealer / supplier
for a new set-up.
Check power cord connections.
Make sure power cord is plugged
into the power outlet. Check power
source. Replace batteries.
Remove load from the scale.
Wait for load to become stable.
Service required.
M500660-210326-Rev005-UM-e