dymax BlueWave 50 Bedieningshandleiding - Pagina 15
Blader online of download pdf Bedieningshandleiding voor {categorie_naam} dymax BlueWave 50. dymax BlueWave 50 20 pagina's. Uv light-curing spot lamp system
TROUBLESHOOTING
WARNING:
Only qualified maintenance personnel should attempt the following procedures:
Seulement personnel d'entretien diplomé devrais essayer les procedures suivant.
AVERTISSEMENT:
Problem
Possible Cause
Improper connections
Bulb Will Not
Bulb beyond useful life of
Ignite
2,000 hours
Main line Fuse blown
(nothing in unit operates)
Bulb beyond useful life
Transmission loss in light
guide too great
Low Output
Intensity
or
Contaminants on
Fails To Cure
Lightguide
Adhesive In
Allotted Time
Bulb/Reflector Assembly
not installed properly
®
BlueWave
50 Light-Curing Spot Lamp
Testing
Visually inspect all input/output
connections (i.e. Power Cord,
Bulb).
Replace with a new Bulb/
Reflector Assembly. Reset the
Bulb Hour Meter and re-test unit.
Remove the Fuse from the
Power Receptacle and check it
with an Ohmmeter.
Use a Radiometer (DYMAX
ACCU-CAL™ 50 or equivalent)
to measure output intensity.
Compare Lightguide output
against new Lightguide (or use
the DYMAX Lightguide
Simulator) to determine
transmission loss.
Visually examine ends of
Lightguide for contaminants.
Visually check to make sure the
Bulb/Reflector Assembly is
seated flush in the Bulb Mount
Assembly (any error in
installation could cause a low
output).
Corrective Action
Secure all connections.
Replace the Bulb/Reflector
Assembly if required
(typical life = 2,000 hours).
Replace the Fuse if defective.
Replace the Bulb/reflector
assembly if beyond useful life
(typical = 2,000 hours).
Replace the Lightguide.
Clean with isopropyl alcohol
(or equivalent). Heavy
deposits on liquid Lightguides
may be removed with a razor
blade. Replace Lightguide if it
can not be cleaned.
Properly install Bulb/Reflector
Assembly.
15