Great Life Healthcare UTS Gebruikershandleiding - Pagina 12
Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Great Life Healthcare UTS. Great Life Healthcare UTS 18 pagina's.
The Universal Therapy System (UTS) is intended for use in the electromagnetic environment (for home and
professional healthcare) specified below.
The customer or the user of the Universal Therapy System (UTS) should assure that it is used in such an
environment.
Immunity test
Electrostatic
Contact: ±8 kV
discharge (ESD) IEC
Air±2 kV, ±4 kV, ±8 kV,
61000-4-2
±15 kV
Electrical fast
+ 2kV for power supply
transient/burst IEC
lines
61000-4-4
+ 1kV for input/output
lines
Surge IEC 61000-4-
+ 0.5kV, +1kV line(s) to
5
line(s)
+ 0.5kV, +1kV, + 2kV
line(s) to earth
Voltage Dips, short
Voltage dips:
interruptions and
0 % UT; 0,5 cycle
voltage variations
0 % UT; 1 cycle
on power supply
70 % UT; 25/30 cycles
input lines IEC
61000-4-11
Voltage interruptions:
0 % UT; 250/300 cycle
Power frequency
30 A/m
(50, 60 Hz)
50 Hz or 60 Hz
magnetic field IEC
61000-4-8
NOTE
UT is the a.c. mains voltage prior to application of the test level.
Manufacturer's declaration-electromagnetic immunity
IEC 60601
test level
12
Compliance level
Contact: ±8 kV
Air±2 kV, ±4 kV, ±8 kV,
±15 kV
+ 2kV for power supply
lines
Not applicable
+ 0.5kV, +1kV line(s) to
line(s)
Not applicable
Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 25/30 cycles
Voltage interruptions:
0 % UT; 250/300 cycle
30 A/m
50 Hz and 60 Hz
Electromagnetic
environment-guidance (for
home and professional
healthcare environment)
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Mains power quality should
be that of a typical home
healthcare environment.
Mains power quality should
be that of a typical home
healthcare environment.
Mains power quality should
be that of a typical home
healthcare environment. If
the user of the Universal
Therapy System (UTS)
requires continued operation
during power mains
interruptions, it is
recommended that the
Universal Therapy System
(UTS) be powered from an
uninterruptible power supply
or a battery.
The Universal Therapy
System (UTS) power
frequency magnetic fields
should be at levels
characteristic of a typical
location in a typical home
healthcare environment.